GREISINGER GMH 3692 Operating Manual Download Page 3

H74.0.14.6C-05 

Operating Manual GMH 3692 

page 3 of 12 

2  Designated Use 

The GMH 3692 is measuring oxygen in air and gas mixtures either as partial pressure or as concentration in 
%vol.  
For the measuring an external sensor of the type GOO-... or GGO... has to be connected to the MINI-DIN-
socket. 
Due to the properties of the sensor, it has to be calibrated regularly (e.g. at fresh air = 20.95%) to get precise 
values. If the sensor is used up, this will be detected during the calibration, the sensor has to be regenerated 
or replaced before continuing with measuring. 

The safety requirements (see below) have to be observed.  
The device must be used only according to its intended purpose and under suitable conditions.  
Use the device carefully and according to its technical data (do not throw it, strike it, …) 
Protect the device from dirt. 

3  Safety 

3.1  Safety signs and symbols 

Warnings are labeled in this document with the followings signs: 

 

DANGER 

        

Caution!

 This symbol warns of imminent danger, death, serious injuries 

and significant damage to property at non-observance. 

 

Attention!

 This symbol warns of possible dangers or dangerous 

situations which can provoke damage to the device or environment at 
non-observance. 

 

Note!

 This symbol point out processes which can indirectly influence 

operation or provoke unforeseen reactions at non-observance. 

3.2  Safety guidelines 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures 
and special safety advises given in this manual will be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected 

to any other climatic conditions than those stated under "Specification". 
If the device is transported from a cold to a warm environment condensation may cause in a failure. In 
such case make sure the device temperature has adjusted to the ambient before trying a new start-up. 

2.    

 

If there is a risk whatsoever involved in running it, the device has to be switched off 
immediately and to be marked accordingly to avoid re-starting. 
Operator safety may be a risk if: 
- there is visible damage to the device 
- the device is not working as specified 
- the device has been stored under unsuitable conditions for a longer time. 
In case of doubt, please return device to manufacturer for repair or maintenance. 

3. 

 

 

 

Do not use these products as safety or emergency stop devices or in any other application 

where failure of the product could result in personal injury or material damage. 
Failure to comply with these instructions could result in death or serious injury and material 
damage. 

 

4. 

 

 

 

This device must not be used at potentially explosive areas! The usage of this device at 

potentially explosive areas increases danger of deflagration, explosion or fire due to 
sparking.  

 

   

 DANGER

 

 

       

 

 

DANGER

 

 
 
 

       

  

 

DANGER

 

 
 
 

Summary of Contents for GMH 3692

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 2 Operating Manual GMH 3692 Oxygen meter for oxygen in gases with integrated temperature and pressure m...

Page 2: ...BRATION OF THE SENSOR 9 7 1 ONE POINT CALIBRATION AL 1 PT 9 7 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 9 7 3 EVALUATION OF SENSOR STATE ELE 9 7 4 CALIBRATION INTERVAL INT 10 8 INSPECTION OF THE ACCURAC...

Page 3: ...esigned and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and spec...

Page 4: ...the real time clock has to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V po...

Page 5: ...Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 Logg arrow No function 7 Warning sign Signals weak battery or other warning message Di...

Page 6: ...points depend on current device settings To change device settings press Menu for 2 seconds This will activate the configuration menu main display Set Pressing Menu changes between the menus points p...

Page 7: ...0 C or 9 0 F 9 0 F The zero point of the temperature measuring is shifted for the entered value This can be used to compensate sensor and instrument deviations OFF No zero adjustment for temperature m...

Page 8: ...life time also suitable for application with larger CO2 concentrations GOEL 380 Fast responding for low oxygen concentration e g protection atmosphere below 1 max 25 For application without larger CO...

Page 9: ...xt reference possible references are blinking in the display 3 Second calibration reference Pt 2 The display will show and the referring reference which should be applied A R for ambient air 0 2 for p...

Page 10: ...cord and display the measuring values GMHKonfig Software for a comfortable configuration of the device e g freeware In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...ue too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery w...

Page 12: ...Break proof ABS housing Protection class Front side IP65 Dimensions L B H mm Weight Without pressure port 142 x 71 x 26 mm L x B x H approx 160 g incl battery Power supply Current consumption 9V Batt...

Reviews: