GREISINGER GMH 3692 Operating Manual Download Page 10

H74.0.14.6C-05 

Operating Manual GMH 3692 

page 10 of 12 

7.4  Calibration Interval (

[.INT

You can enter the interval after which the device reminds you to recalibrate in the configuration.  
The interval times should be chosen according to the application and the stability of the sensor.  
“CAL” flashes on the display as soon as the interval has expired.  

8  Inspection of the accuracy / Adjustment Services 

The instrument can be sent to the manufacturer for adjustment and function test. 
Only the manufacturer can check all systems on correct them if necessary. 

Calibration certificates 

– DKD-certificates – other certificates: 

If device should be certificated for its accuracy, it is the best solution to return it to the manufacturer. 
(please specify references, e.g. 20.9 and 100%). 
If the instrument is certified with its sensor, this proves for example the linearity of the measuring chain, 
regular recalibration by the user is still necessary! 

9  Serial Interface 

With an electrically isolated interface converter USB3100, GRS3100 or GRS3105 (accessory) the device can 
be connected to a PC. 
With the GRS3105 it is possible to connect up to 5 instruments of the GMH3000 family to a single interface 
(please also refer to GRS3105-manual). As a precondition the base addresses of all devices must not be 
identical, make sure to configure the base addresses 

accordingly (refer menu point “Adr.” in chapter 5).  

In order to avoid transmission errors, there are several security checks implemented (e.g. CRC). 

The following standard software packages are available for data transfer: 

 

GSOFT3050

:  

Operating and evaluation software for instruments with integrated logger function 

 

EBS20M/ -60M

20- / 60-channel software to record and display the measuring values 

 

GMHKonfig

Software for a comfortable configuration of the device (e.g. freeware) 

In case you want to develop your own software we offer a 

GMH3000-development package

 including 

an universally applicable 32bit Windows functions library ('GMH3000.DLL') with documentation that can 
be used by all 'serious' programming languages. 

Programming examples for Visual Studio 2010 (C#, C++), 

Visual Basic 6.0™, Delphi 1.0™, Testpoint™, 

Labview™ 

The Device has 4 Channels: 

- oxygen concentration % Vol 
- oxygen partial pressure in hPa or mmHg 
- temperature value at the time of recording in °C or °F 
- absolute pressure in hPa abs or mmHg abs 

Supported Interface-functions:  

1  2  3  4  Code 

Name/Function 

1  2  3  4  Code 

Name/Function 

x  x  x  x  0 

read nominal value 

x  x  x  x  199 

read measuring type in display 

x  x  x  x  3 

read system status 

x  x  x  x  200 

read min. display range 

x   

 

 

12 

read ID-no. 

x  x  x  x  201 

read max. display range 

x  x  x   

22 

read min alarm limit 

x  x  x  x  202 

read unit of display 

x  x  x   

23 

read max alarm limit 

x  x  x  x  204 

read decimal point of display 

x  x  x  x  176 

read min measuring range 

x   

 

 

208 

read channel count 

x  x  x  x  177 

read max measuring range 

x   

 

 

222 

read turn-off-delay 

x  x  x  x  178 

read measuring range unit 

x   

 

 

223 

Set turn-off-delay 

x  x  x  x  179 

read measuring range decimal point 

x   

 

 

240 

Reset 

x  x  x  x  180 

read measuring type  

x   

 

 

254 

read program identification 

The measuring and range values read via interface are always in the selected display unit! 

 

10 

Alarm („AL.“) 

There are three possible settings: 
Alarm off (AL. oFF), on with buzzer (AL. on), on without buzzer (AL. no.So). 
Following conditions will display an alarm, when the function is activated (on or no.So): 

-

 

Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi). 

-

 

Sensor error  

-

 

Low battery (bAt) 

-

 

Err.7: System error (always with buzzer) 

In 

case of an alarm and when polling the interface the “prio”-flag is set in the returned message. 

 

Summary of Contents for GMH 3692

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 2 Operating Manual GMH 3692 Oxygen meter for oxygen in gases with integrated temperature and pressure m...

Page 2: ...BRATION OF THE SENSOR 9 7 1 ONE POINT CALIBRATION AL 1 PT 9 7 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 9 7 3 EVALUATION OF SENSOR STATE ELE 9 7 4 CALIBRATION INTERVAL INT 10 8 INSPECTION OF THE ACCURAC...

Page 3: ...esigned and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and spec...

Page 4: ...the real time clock has to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V po...

Page 5: ...Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 Logg arrow No function 7 Warning sign Signals weak battery or other warning message Di...

Page 6: ...points depend on current device settings To change device settings press Menu for 2 seconds This will activate the configuration menu main display Set Pressing Menu changes between the menus points p...

Page 7: ...0 C or 9 0 F 9 0 F The zero point of the temperature measuring is shifted for the entered value This can be used to compensate sensor and instrument deviations OFF No zero adjustment for temperature m...

Page 8: ...life time also suitable for application with larger CO2 concentrations GOEL 380 Fast responding for low oxygen concentration e g protection atmosphere below 1 max 25 For application without larger CO...

Page 9: ...xt reference possible references are blinking in the display 3 Second calibration reference Pt 2 The display will show and the referring reference which should be applied A R for ambient air 0 2 for p...

Page 10: ...cord and display the measuring values GMHKonfig Software for a comfortable configuration of the device e g freeware In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...ue too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery w...

Page 12: ...Break proof ABS housing Protection class Front side IP65 Dimensions L B H mm Weight Without pressure port 142 x 71 x 26 mm L x B x H approx 160 g incl battery Power supply Current consumption 9V Batt...

Reviews: