GREISINGER GMH 3692 Operating Manual Download Page 11

H74.0.14.6C-05 

Operating Manual GMH 3692 

page 11 of 12 

11 Error and System Messages 

Display 

Meaning 

Remedy 

 

low battery voltage, device will continue to work 
for a short time 

replace battery 

If mains operation: wrong voltage 

replace power supply, if fault continues to 
exist: device damaged 

 

low battery voltage 

replace battery 

If mains operation: wrong voltage 

Check/replace power supply, if fault 
continues to exist: device damaged 

No display 

or 

weird display 

 

Device does not 

react on keys 

low battery voltage 

replace battery 

If mains operation: wrong voltage 

check/replace power supply, if fault 
continues to exist: device damaged 

system error 

disconnect battery or power supply, wait 
some time, re-connect 

device defective 

return to manufacturer for repair 

SENS 

ERRO

 

sensor error: no sensor cable connected 

connect suitable sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.1 

Value exceeding measuring range 

Check: Is the value exceeding the 
measuring range specified? ->value too 
high! 

Wrong sensor connected 

Check sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.2 

Value below display range 

Check: Is the value below the measuring 
range specified? ->value too low! 

Wrong sensor connected 

Check sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.7 

system error 

return to manufacturer for repair 

If 

BAT

 is flashing, the battery will be exhausted soon. Further measurements are possible for short time. 

If 

BAT

 is displayed continuously the battery is ultimately exhausted and has to be replaced. Further 

measure

ments aren’t possible any more.

 

Messages During Calibration/Adjustment 

>CAL< 

CAL flashing in 

display 

either preset calibration interval has expired or 
last calibration is not valid 

device has to be calibrated! 

[AL  ERR.1

 

wrong reference point at air  

check sensor and reference gas 

[AL  ERR.2

 

slope too low 

 

reference gas wrong 

check sensor and reference gas 

sensor element is defect  

replace sensor element  

[AL  ERR.3

 

slope too high 

 

reference gas wrong 

check sensor and reference gas 

s

ensor element is defect  

replace sensor element  

[AL  ERR.4

 

incorrect calibration temperature 

calibration can only be done at 0…50 °C 

[AL  ERR.5

 

Zero value to low/negative 

 

sensor element is defect  

replace sensor element  

[AL  ERR.6

 

zero value to high 

 

reference gas wrong 

check sensor and reference gas 

s

ensor element is defect  

replace sensor element  

[AL  ERR.7

 

incorrect calibration pressure 

check calibration pressure 

[AL  ERR.8

 

signal not stable / timeout 

check sensor and reference gas 

[AL  ERR.9

 

sensor not known: cannot be calibrated 

check sensor and wiring 

 

 

Summary of Contents for GMH 3692

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 2 Operating Manual GMH 3692 Oxygen meter for oxygen in gases with integrated temperature and pressure m...

Page 2: ...BRATION OF THE SENSOR 9 7 1 ONE POINT CALIBRATION AL 1 PT 9 7 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 9 7 3 EVALUATION OF SENSOR STATE ELE 9 7 4 CALIBRATION INTERVAL INT 10 8 INSPECTION OF THE ACCURAC...

Page 3: ...esigned and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and spec...

Page 4: ...the real time clock has to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V po...

Page 5: ...Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 Logg arrow No function 7 Warning sign Signals weak battery or other warning message Di...

Page 6: ...points depend on current device settings To change device settings press Menu for 2 seconds This will activate the configuration menu main display Set Pressing Menu changes between the menus points p...

Page 7: ...0 C or 9 0 F 9 0 F The zero point of the temperature measuring is shifted for the entered value This can be used to compensate sensor and instrument deviations OFF No zero adjustment for temperature m...

Page 8: ...life time also suitable for application with larger CO2 concentrations GOEL 380 Fast responding for low oxygen concentration e g protection atmosphere below 1 max 25 For application without larger CO...

Page 9: ...xt reference possible references are blinking in the display 3 Second calibration reference Pt 2 The display will show and the referring reference which should be applied A R for ambient air 0 2 for p...

Page 10: ...cord and display the measuring values GMHKonfig Software for a comfortable configuration of the device e g freeware In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...ue too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery w...

Page 12: ...Break proof ABS housing Protection class Front side IP65 Dimensions L B H mm Weight Without pressure port 142 x 71 x 26 mm L x B x H approx 160 g incl battery Power supply Current consumption 9V Batt...