4 | Operation
G 1690 / G 1690T
10 / 20
B-H86.0.31.DK2-1.0
4
Operation
4.1
Opening the configuration menu
1.
Press the
Function button
for 2 seconds to open the
Configuration
menu.
2.
(ONF
appears in the display. Release the
Function button
.
Parameter
Values
Meaning
__
INP
Auxiliary display
(only available at G 1690T)
%
02
No auxiliary display,
only display of the oxygen concentration in [vol. %]
m
80D
Auxiliary display:
maximum operational depth MOD in meter [m] sea
water
(for explanation of MOD see #1).
FT
80D
Auxiliary display:
maximum operational depth MOD in feet [ft] sea
water
(for explanation of MOD see #1).
__
P.ABS
Air pressureö
500 ... 1 100
Air pressure in [hPa] corresponds to [mbar]
(note explanation #2 )
__
PO2
Max. Oxygen partial pressure
(only available at G 1690T)
0.2 ... 1.6
Partial pressure in [bar]
(Observe extended notes #3 !!)
POFF
Shut-off time
OFF
No automatic shut-off
0:15
,
0:30
,
1:00
,
4:00
,
12:00
Automatic shut-off after a selected time in
hours:minutes, during which no buttons have been
pressed
=
=