G 1400 series
Operation | 4
B-H87.0.0X.DK2-3.0
9 / 22
4
Operation
4.1
Opening the configuration menu
1.
Press the
Function key
for 2 seconds to open the
Configuration
menu.
2.
(ONF
appears in the display. Release the
Function key
.
_
Parameter
Values
Meaning
__
INP
Measuring unit
(OND
conductivity
RES,
specific resistance
(G 1420 only)
E(
EC (~ mS/cm)
(G 1409 only)
(F
CF (~ 10 x EC)
(G 1409 only)
SAL
salt content / salinity
(G 1410 only)
TDS
total dissolved solids
(G 1409, G 1410 only)
__
CTDS
Factor for TDS
(G 1410 and G 1409 only)
0
.
40
..
1
.
00
Conversion factor for TDS measurement
commonly used: 0.500 or 0.700
__
TCOR
Temperature compensation
OFF
Do not compensate conductivity measurement
NLF
Non-linear function for natural water in accordance
with EN 27888 (ISO 7888) Ground water, surface
water or drinking water
NACI
Compensation of weak NaCI solutions only in pure
and ultrapure water (G 1420 only)
L,N
Linear temperature compensation (G 1420 only)