G 1400 series
Operation | 4
B-H87.0.0X.DK2-3.0
11 / 22
4.2
Adjustment of the measuring input
With the conductivity slope correction, the conductivity value can be readjusted.
The temperature input can be adjusted with the zero point correction and the gradient
correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signalled with the display text
T.OF
,
T.SL
or
S(L
when switching on.
1.
Switch the device off.
2.
Hold the
down button
and press the
On/Off button
briefly to switch on the device
and open the
Adjustment
menu.
3.
The display shows the first parameter. Release the
down button
.
Parameter
Values
Meaning
_
T.OF
Zero point correction of the temperature
0.00
No zero point correction
-5.00
…
5.00
Zero point correction in °C. (at °F -9.00 .. +9.00)
_
T.SL
Gradient correction of the temperature
0.00
No gradient correction
-5.00
…
5.00
Gradient correction in %
_
S(L
Gradient correction for the conductivity value
1.000
No gradient correction
0.800
…
1.200
Multiplier for the gradient correction
Formula used by device:
Temperature = °C: Display = (measured value -
T.OF
) * (1 +
T.SL
/ 100)
Temperature = °F: Display = (meas. value - 32 °F -
T.OF
) * (1 +
T.SL
/ 100) + 32 °F
Gradient correction conductivity.: Displayed value = measured value /
S(L