![GREISINGER G 1400 Series Quick Reference Manual Download Page 21](http://html1.mh-extra.com/html/greisinger/g-1400-series/g-1400-series_quick-reference-manual_2264316021.webp)
G 1400 series
Technical data | 8
B-H87.0.0X.DK2-3.0
21 / 22
Standard functions
Min/Max/Hold
Adjustment
Offset and gradient correction for temperature,
Gradient correction - conductivity
Housing
Break-proof ABS housing
Protection rating
IP65 / IP67
Dimensions L*W*H 108 * 54 * 28 mm, without measuring cell or kink protection
Weight
165 g incl. battery and measuring cell (G 1409)
180 g incl. battery and measuring cell (G 1410)
210 g incl. battery and measuring cell (G 1420)
Nominal temperature
25 °C
Operating conditions
-20 to 50 °C; 0 to 95 %RH (temporarily condensing)
Storage temperature
-20 to 70 °C
Current supply
2 * AA batteries (mignon)
Current requirement approx. 2.2 mA, approx. 3.5 mA with lighting
battery life
Service life > 1000 hours with alkaline batteries
(without backlighting)
Battery indicator
4-stage battery status indicator,
Replacement indicator for depleted batteries: "BAT"
Auto-power-OFF
function
The device switches off automatically if this is activated
Directives and
standards
The devices conform to the following Directives of the
Council for the harmonisation of legal regulations of the
Member States:
2014/30/EU EMC Directive
2011/65/EU RoHS
Applied harmonised standards:
EN 61326-1:2013
Emission limits: Class B
Immunity according to Table 1
Additional errors: < 1 % FS
EN IEC 63000:2018
The device is intended for mobile use and/or stationary
operation in the scope of the specified operating conditions
without further limitations.