background image

H63.0.21.6C-01 

Operating Manual GMR 110 

page 6 of 12 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

5.3  Temperature compensation  

The temperature compensation is important for a reliable moisture-measuring.  
There for the device features a automatic temperature compensation (Atc) or manual temperature 
compensation input. The choice between the two possibilities is done in the Menu (Atc on/off) 
According to the selected material characteristic curve the device will use the associated tempera-
ture compensation. 

Calling temperature when using automatig temperature compensation (Atc on): 

  

• press shortly 

:  in the display will appear: 

t.°C

 or 

t.°F 

and the temperature 

  

• press 

 again to switch back to the measuring display. 

 

Temperature selection: 

  

• press shortly 

:  in the display will appear: 

t.°C

 or 

t.°F 

and the temperature 

or  .

 

  

• set the temperature by pressing  

  

• press 

 again to switch back to the measuring display. 

5.4  Auto-Hold function 

Particularly when measuring dry materials, electrostatic charges and other similar noise could dither 
the measuring value. With activated auto-hold function the device will acquire an exact measuring 
value automatically. During that, the device could be put down to avoid noise through discharge of 
the clothing etc.  
After having acquired the measuring value, the display will change to „HLD‟: The value will be fro-
zen as long as a new measuring is initiated by pressing key 3 (hold). 

5.5  Measuring in wood 

For measuring wood, punch in the measuring-pikes across to the wood-grain, having a good con-
tact between the pikes and the wood (measuring along wood-grain deviates minimal). 

DO NOT HIT ONTO THE DEVICE OR PUNCH THE NEEDLES IN PEPPY 

The device may be damaged. 
Select

 correct wood-sort

 (see chapter 4.4 or Appendix A). 

Ensure to have entered the 

correct temperature

 (see chapter 5.3).  

Now read the measuring-value or when having activated the auto-hold-function initiate a new 
measuring by pressing 

hold

 (key 3) . 

The measured resistance will be extremely high when measuring dry wood (<15%) thus the meas-
uring will need more time to achieve its terminal value among other things static discharge could 
momentarily falsify the measuring. Therefore beware of static discharge and wait long enough until 
a stable measuring value is displayed (unstable

: „%“ blinking) or use the auto-hold-function (see 

chapter 5.4 Auto-Hold function). 
When measuring very wet wood (e.g.>50%u) the measuring value may suffer from polarisation ef-
fects (steadily decreasing measuring value). In this case the value 5 seconds after the insertion of 
the electrodes respectively after switching on the instrument is valid.  

Very accurate measures can be carried out within the range of 

6 to 30%

.  

Beyond this range the accuracy will lessen, but the device will deliver reference values still sufficient 
for the practitioner. 
It is measured between the measuring-pikes insulated among each other. Requirements for an ex-
act measurement: 

-

  choose correct place to measure: place should be free of irregularities like resin

–clusters, knurls, 

rifts, etc.  

-

  choose correct measure depth: Recommendation: for trimmed timber: punch in the pikes up to 

1/3 of the material thickness. 

Summary of Contents for GMR 110

Page 1: ...93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de as of version 2 1 GMR 110 Operating Manual Resistive material moisture measuring instrument with integrated...

Page 2: ...ON OF THE DEVICE 9 7 ERROR AND SYSTEM MESSAGES 9 8 RESHIPMENT AND DISPOSAL 10 9 SPECIFICATION 10 10 APPENDIX A ADDITIONAL WOOD SORTS 11 1 General Note Read this document carefully and get used to the...

Page 3: ...ot be guaranteed unless the stan dard safety measures and special safety advises given in this manual will be adhered to when using the device 1 Trouble free operation and reliability of the device ca...

Page 4: ...th above specification do not throw hit against etc Protect from soiling 4 Handling 4 1 Display elements 1 Main display Display of the current moisture or water content 2 Material display The name of...

Page 5: ...factors like the ap plication field of the measured material have to be taken into account for the final rating Experience and knowledge can only be supported by this instrument not replaced 5 2 Moist...

Page 6: ...r measuring wood punch in the measuring pikes across to the wood grain having a good con tact between the pikes and the wood measuring along wood grain deviates minimal DO NOT HIT ONTO THE DEVICE OR P...

Page 7: ...ies of plasters and the influence dam age by moisture the precision of the measuring is not as high as the precision when measuring in wood However precise measuring values aren t necessary in the mos...

Page 8: ...mpound insert profuse conductivity compound between the holes and the brush type probe and let the electrode stick in the hole for about 30min before measuring with the device switched off Tempera tur...

Page 9: ...ntent w Display unit of temperature input factory setting C Temperature input in C F Temperature input in F AtC Automatic temperature compensation AtC factory setting on 0F AtC deactivated manual inpu...

Page 10: ...or wood an construction materials Measuring ranges 0 0 50 w water content 0 0 100 u material moisture Resolution 0 1 over 19 9 1 for u and w Moisture rating 6 stage bar graph display from WET to DRY A...

Page 11: ...ercus petraea h 01 chestnut sweet Castanea sativa h 01 obeche Triplochiton scleroxylon h 02 common beech Fagus sylvatica h 02 okoum Aucoumea klaineana h 02 cypress Cupressus spp omu Entandrophragma ca...

Page 12: ...H63 0 21 6C 01 Operating Manual GMR 110 page 12 of 12 _____________________________________________________ _____________________________________________________________________________...