45CA-11 and 45CA-12
RA0403/RA0404 Hi-Res Ear Simulator
Standards, based on:
IEC 60318-4 (compatible)
Occluded-ear simulators for the measurement of earphones coupled to the ear by
ear inserts.
ITU-T Recommendation P.57
Series P: Telephone transmission quality, Objective measuring apparatus: Artificial ears
ANSI S3.25/ASA-2009
American National Standard For an Occluded Ear Simulator
Frequency Response
Typical transfer impedance
re 500 Hz
Resonance frequency
13.5 kHz ± 1 kHz
Sensitivity
Sensitivity
1.6 mV/Pa
Dynamic Range
RA0403 (LEMO)
44 dBA to 169 dB
RA0404 (CCP)
46 dBA to 174 dB
Effective Volume
at 500 Hz
1260 mm
3
Dimensions
Height
23.0 mm
Diameter
23.75 mm
-5
0
5
10
15
20
25
30
35
40
45
50
55
100
1000
10000
100000
d
B
r
e
5
00
H
z
Frequency [Hz]
Typical response RA0403
GRAS RA0403 Tolerance
IEC Tolerance
Instruction Manual