45CA-3 and 45CA-4
RA0039 Ear Simulator
Standards
IEC 60318-1
Ear simulator for the calibration of supra-aural and circumaural earphones
ITU-T Recommendation P.57
Series P: Telephone transmission quality, Objective measuring apparatus: Artificial
ears
Frequency Response
Typical frequency response
Sensitivity
Sensitivity
12.5 mV/Pa
Dynamic Range
with 40AG
20 dBA to 164 dB
with 40AO
20 dBA to 163 dB
Dimensions
Height
19.8.0 mm
Diameter
60 mm
Environmental Calibration Conditions
Temperature
23 °C ±3 °C
Relative humidity
60 % ±20 %
Barometric pressure
101.3 kPa ±3 kPa
Instruction Manual
3 0 | GRAS 45CA Headphone/Hearing-protector Test Fixture