45CA-7 to -12
As described on page 22 the anthropometric pinna is mounted directly on the Ear Simulator,
without the use of en ear canal extension. Therefore, for calibration with a pistonphone with a
1/2” coupler, the ear simulator and the anthropometric pinna must be removed to allow mounting
of the GR0408 External Ear Canal.
Removing the Ear Simulator and Anthropometric Pinna
1.
Loosen the screw holding the ear simulator, see Fig. 7 on page
22.
2.
Gently pull out the assembly. See also Fig. 8 on page
23.
Important.
Do NOT pull at the ear lobe.
GR0917
GR0924
GR1440B
RA0045
26A
S
45EA
46BP
RA0241
40A
G
GS0158
GR0010
46BD
1
2 4
3 5
L
B-2
B-1
0
20
10
01
02
0
J
GR1153
46BP
-1
40A
D
GR0917
GR0924
GR1440B
RA0045
26A
S
45EA
46BP
RA0241
40A
G
GS0158
GR0010
46BD
1
2 4
3 5
L
B-2
B-1
0
20
10
01
02
0
J
GR1153
46BP
-1
40A
D
Fig. 13.
Removing the anthropometric pinna from the ear simulator.
3.
Unscrew the pinna simulator from the ear simulator
4.
Mount the external ear canal with the union nut onto the ear simulator.
GR0917
GR0924
GR1440B
RA0045
26A
S
45EA
46BP
RA0241
40A
G
GS0158
GR0010
46BD
1
2
4
3
5
L
B-2
B-1
0
20
10
01
02
0
J
GR1153
46BP
-1
40A
D
GR0917
GR0924
GR1440B
RA0045
26A
S
45EA
46BP
RA0241
40A
G
GS0158
GR0010
46BD
1
2
4
3
5
L
B-2
B-1
0
20
10
01
02
0
J
GR1153
46BP
-1
40A
D
GR0409
Union Nut
GR0408
External Ear Canal
RA0045/-S1 or RA0401/2
Ear Simulator
Fig. 14.
Showing the External Ear Canal and Union Nut. For clarity, preamplifier and cable are not shown.
You are now ready to calibrate. The Ear Simulator fits directly into a pistonphone with a 1/2” cou
-
pler. Refer to your pistonphone manual for further information. See the corrections factors in the
table below.
Instruction Manual
45CA Headphone/Hearing-protector Test Fixture | 2 7