Connect the output side of the AC adapter to the
connector indicated as "DC LINE" on the GL220.
Use a flathead screwdriver to push the button above the
ground terminal while connecting the grounding cable to the
GL220. Connect the other end of the cable to ground.
Grounding cable
Note: Connect wires to the desired terminals according to the terminal numbers on
top of the terminal block.
Direct voltage input
Thermocouple input
Direct current input
CH 1 2 3 4 5 6 7 8 9 10
Shunt resistance
Ex: for current in the 4 to 20 mA
range, apply a resistance of 250 W
(±0.1%) and perform measurement
in the 1 to 5 V ra
nge.
Note: Use B-551 (option)
for this shunt resistance.
+
-
Compensation
copper wire
+
-
Direct voltage
+
-
Direct voltage
*B-513 (sold separately) cable is required for external input/output.
(For logic/pulse input, alarm output, trigger input, external sampling input
B-513
Orange with red dotted line : 1
Orange with black dotted line : 2
Grey with red dotted line : 3
Grey with black dotted line : 4
White with red dotted line : 1
White with black dotted line : 2
Yellow with red dotted line : 3
Yellow with black dotted line : 4
Pink with red dotted line : Trigger input/
external sampling input
Shielded
GND
Logic/pulse
input
Alarm output
Connection diagram
Pink with black dotted line
GL220 Connection Procedures
Connecting the AC Adapter
Connecting the Grounding Cable
Making Connections to the Analog Input Terminals
3
Making Connections to the External Input/Output Terminals