![GRAPHTEC GL350 Service Manual Download Page 32](http://html1.mh-extra.com/html/graphtec/gl350/gl350_service-manual_2257291032.webp)
GL400-UM-251-9370
4-9
4. SETUP PROCEDURES
Setup procedure
(1) Connect the resistor box and the GL450/400/350 for the 0-ohm setting as shown above.
(2) Leave the GL450/400/350 turned on for at least
10 minutes
.
(3) After entering the System Setup mode, open the Setup menu as shown below.
Specify the AMP Setup conditions as shown below.
Mode
:Setup3 (RTD temp.)
Slot
:Slot1
CH
:CH 2
(4) Move the cursor to the “Execute Setup” position as shown above.
(5) Press the [ENTER] key to start the Setup3 (RTD temp.) operation.
Message
(6) The “
0 ohm : [START]
” message is displayed at the location shown above.
(7) Input “0 ohm” to the input terminals.
(8) Press the [START] key and then wait until the "
350 ohm : [START]
" message appears.
(9) Connect the resistor box and the GL450/400/350 for the 350-ohm setting as shown above.