GL400-UM-251-9370
4-4
4. SETUP PROCEDURES
(8) Press the [START] key and then wait until the "
20 mV : [START]
" message appears.
Specify the voltage generator's output voltage setting to correspond to one of the displayed voltage
ranges in the following table, and then input that voltage to the GL450/400/350.
Message
Input Voltage
GND
0.000 mV
20 mV
20.000 mV
50 mV
50.000 mV
100 mV
100.000 mV
200 mV
200.000 mV
500 mV
500.000 mV
1 V
1.000 V
2 V
2.000 V
5 V
5.000 V
10 V
10.000 V
20 V
20.000 V
50 V
50.000 V
(9) Repeat step (8) by sequentially setting the Measure parameter to 50 mV, 100 mV, 200 mV, 500 mV, 2 V,
5 V, 10 V, 20 V and 50 V, and then inputting the corresponding voltage.
(10) When you have finished the Setup operation for 50 V, the "
EEPROM : [ENTER]
" message appears.
Press the [ENTER] key to begin writing the new Setup data to the EEPROM.
(11) After the setup parameters have been registered to the EEPROM, the display closes and returns to
normal mode.
(12) Turn off the GL450/400/350.
This completes the voltage amp setup.