4. SETUP PROCEDURES
(10) Re-open the Setup menu as shown below.
Specify the AMP Setup conditions as shown below.
Mode
:Setup2 (Room temp.)
Slot
:Slot1
Temperature Sensor :Sensor1
Area
:Area2
CH
:CH 3
(11) Repeat steps (4) to (8) for Area2 of Sensor1.
(12) If you have the 20-channel input terminal unit, repeat steps (4) to (8) for Sensor2.
(13) If you have the 50-channel input terminal unit, repeat steps (4) to (8) until Area 5 of Sensor2.
(14) Turn off the GL450/400/350.
This completes the Room temp. setup.
Setup flow for the Room temp.
Procedure
10-channel input terminal
20-channel input terminal
50-channel input terminal
1
Sensor 1
Area 1
Sensor 1
Area 1
Sensor 1
Area 1
2
Sensor 1
Area 2
Sensor 1
Area 2
Sensor 1
Area 2
3
None
Sensor 2
Area 1
Sensor 1
Area 3
4
None
Sensor 2
Area 2
Sensor 1
Area 4
5
None
None
Sensor 1
Area 5
6
None
None
Sensor 2
Area 1
7
None
None
Sensor 2
Area 2
8
None
None
Sensor 2
Area 3
9
None
None
Sensor 2
Area 4
10
None
None
Sensor 2
Area 5