Overhead Stirrer Multi Mixer MM-1000
Edition 2.04
– December of 2021
Operating Manual
Page 7
6. Program setting
6.1.
The program consists of cycles. Each cycle includes motions of three different types (rotational, re-
ciprocating and vibro) set one after another with the duration from 0 to 250 seconds for rotational and
reciprocal motion types and from 0 to 5 seconds for vibro motion.
6.2.
Changing parameters.
Press the
Select
key (fig. 2/9) to choose the parameter to change, each
consecutive press activates the next parameter in cycle. Active parameter is flashing. Use the ▲ and
▼ keys (fig. 2/10) to set the necessary value. Pressing the key for more than 2 s increases value
changing speed.
6.3.
All parameters except overall timer can also be changed during the operation. Unit automatically
saves and applies changes to the current work program.
6.4.
Following parameters are available: rotation and reciprocation speed, reciprocation and vibro turning
angle, time for each motion type and overall operation timer.
6.5.
Skip a motion.
If the time for a type of motion is set to zero (indication OFF), this type of motion will
be skipped in the cycle.
6.6.
Pause.
If the turning angle for reciprocating or vibro motions is set to zero (indication 0), unit will
interpret this as a pause. The stirring element will not move in this mode during the operation and the
overall timer will be counted down.
6.7.
The overall timer (fig. 2/7) is used to control the operation time. The timer can be set for the period
from 1 min to 96 hours (increment 1 min). If the timer is set to 00:00, unit operates non-stop, see
6.8.
Table 1 shows possible motion combinations in the cycle.
Table 1. Motion type combinations
#
Rotational
Reciprocal
Vibro
#
Rotational
Reciprocal
Vibro
1
ON
ON
ON
8
ON
ON
OFF
2
ON
OFF
ON
9
ON
ON
Pause
3
ON
Pause
ON
10
OFF
ON
ON
4
ON
OFF
OFF
11
OFF
Pause
ON
5
ON
Pause
OFF
12
OFF
ON
Pause
6
ON
OFF
Pause
13
OFF
OFF
ON
7
ON
Pause
Pause
14
OFF
ON
OFF
6.9.
Further examples illustrate program setting for both separate motions and their combinations in the
cycle.
6.9.1.
Rotational motion.
Set the speed (40-1000 rpm, fig. 2/A) and time (1-250 s, fig. 2/B) of rotational
motion. Turn off reciprocal and vibro motions by setting their motion times to zero (indications OFF,
fig. 2/C and 2/D).
MM-1000
is programmed to change the direction of rotation when the motion timer is restarted,
i.e. if the rotation time is set to 30 s, then the direction is changed every 30 s as shown on figure
3, where positive RPM is clockwise motion and negative - counter clockwise.
Figure 2.
Figure 3.