32
14.1.
Double range / interval configuration.
Xtrem scale can be configured as a 2-ranges or 2-interval instrument using
the settings on the
menú.
Select the function mode:
(the scale will work as a single
range instrument),
(the scale will work as a 2 ranges
instrument,
(the scale will work as a 2 interval
instrument).
Second range / interval maximum capacity Max
2
.
Division for the second range / interval: e
2
scale interval;
Possible values are 1, 2, 5, 10, 20 or 50
14.2.
Zero options menu
Configuration settings for the automatic zero setting of the scale after power
on and the scale zero-tracking device.
Activate / deactivate the initial zero setting feature. When
select
, the indication is automatically set to zero when
the weighing instrument is switched on, before any weighing.
This feature is activated as default factory setting, select
to deactivate the initial zero setting of the scale.
Initial Zero Setting Range: Enter the maximum range for the
initial zero setting of the scale as a percent of the maximum
capacity.
Zero-tracking device on / off. As factory default is set to ON.
Maximum effect of the zero-tracking device expressed as
fractions of the scale interval.