15
6.4.
ZERO SETTING
A short press on this key sets the scale to zero. The scale is
deemed to be “set to zero” when the weight on the load receptor
is lower than ¼ of division.
While the scale is “set to zero”, the reading
is shown in the display.
When the scale is set to zero
the automatic “zero tracking” device is in
operation. This function automatically sets to zero when variations of less
than ¼ division occur if they do not add up to more than ¼ division during
one second. This function can be deactivated in the
option of the
configuration menu.
Pressing on the zero key also deactivates the tare if it happens to be
activated.
The zeroing of the scale is limited to 4% of its maximum capacity. Pressing
the zero-setting button will not take effect when this margin is exceeded.
6.5.
HIGH-RESOLUTION MODE
When pressing HR key,
the scale’s resolution is enlarged by x10,
allowing the weight to be viewed with a 10-times-smaller division
for 4 seconds.
An additional digit appears in the weight display, and the decimal point is
shifted to the left by one position. This additional digit is indicated with a
straight angle in its top left corner:
1 Example for a scale with a 1-kg division in normal mode and in high-resolution mode