SCALE SETUP AND CALIBRATION
At the
menu you view and configure the settings needed to define and
adjust the measurement scale of the instrument.
Access to these configuration options is reserved for technical
personnel and is protected by a keyword to avoid accidental changes
that would cause the instrument to malfunction. When trying to
enter the
menu, the terminal will prompt for a password.
Enter pin
“
1009
”
to access the
configuration.
When the weighing scale has been sealed to meet legal regulations,
K3X terminal will not allow changes on most of these settings and will
show the message
when trying to modify them.
The Xtrem unit records the date and time of the last modification in
any of these parameters. Modifying these parameters may imply the
loss of the equipment warranty.
Measurement unit: g, kg
Maximum capacity (Max
1
when 2 ranges or 2 intervals). Enter
the value, including the decimal digits.
Division: Scale interval (e
1
when 2 ranges); the smallest
increment that the instrument can measure. Possible values
are 1, 2, 5, 10, 20 or 50.
Position of decimal point.
Configuration menu for double range or double interval
instrument.
Configuration menu for the instrument’s options associated
to
the automatic initial zero setting.
Test mode that allows to display the load cell signal instead the
weight value. The display will show A/D converter counts.
Scale calibration options.