background image

5

OPERATION

Turning on

Turn on the power, the indicator performs self-checking and go into weighing mode.

Zero function

Press 

 for zero return, indicator returns to zero within the zero range 2%FS. 

Please make sure the stabilization light is on when zero operation.

If obstruction occurs by more than 1 wireless indicator working at same time, 
you can open both indicator and scale, change the number of transcoder 
inside with same number and restart. 

Tare

At weighing mode, press 

 to deduct displaying weight while it is positive 

and stabilization light is on. At no-load status of the scale, press 

  to 

deduct tare and then tare light is off. 

Manual accumulation

At weighing mode, press 

, it displays “ADD-01”“01” means times of 

accumulation (the Max times is 99 and then it will return to 01). After then 
it will display the value of accumulation for 3s and then back to normal 
weighing mode by itself.

Summary of Contents for C6-10T

Page 1: ...SERIES C6 3T 5T 10T 15T 20T OPERATION MANUAL ...

Page 2: ... Front and back view of the indicator 4 Keys function 4 Operation 5 Turning on 5 Zero function 5 Tare 5 Manual accumulation 5 Check accumulation 6 Clear accumulation 6 Parameters setup 6 Calibration 8 Error indication 9 Rechargeable lithium battery 10 Maintenance 10 Guarantee 12 ...

Page 3: ... times second Display Display 0 999999 5 digits LCD LED 7 indicators Division Value 1 2 5 10 20 50 optional Ambient Environment Distance between scale and indicator 150m Power Supply DC 3 7V Ambient Temperature 0 40 ºC Storing Temperature 25 ºC 55 ºC Ambient Moisture 85 RH Warm up Time 10 15 分钟 ...

Page 4: ... than 5s it enters setup mode All the setting in this mode should be recorded with the calibration header on or the data will lose Tare function in weighing status Zero function in weighing status For selecting Weighing accumulation Long press for 1sec to turn on Long Press for 3sec to turn off ...

Page 5: ... open both indicator and scale change the number of transcoder inside with same number and restart Tare At weighing mode press to deduct displaying weight while it is positive and stabilization light is on At no load status of the scale press to deduct tare and then tare light is off Manual accumulation At weighing mode press it displays ADD 01 01 means times of accumulation the Max times is 99 an...

Page 6: ...he current accumulation PARAMETERS SETUP In weighing status hold for more than 5s and entering setup mode P mode all the setting in this mode should be recorded with the calibration header on or the data will lose P1 P13 parameters are available press to switch and press to select parameter Detail for parameter 1 P1 x kg Lb shift X 1 kg mode default X 2 Lb mode 2 P2 x Not available for this model ...

Page 7: ...e X 1 No power saving X 2 Power saving available X 3 None 7 P7 x Zero tracking Scope X 1 0 5e X 2 1 0e X 3 1 5e default X 4 2 0e X 5 2 5e X 6 3 0e X 7 3 5e 8 P8 x Zero key Scope X 1 2 FS X 2 4 FS X 3 10 FS default X 4 20 FS 9 P9 x Zero scope upon start X 1 2 FS X 2 4 FS X 3 10 FS default X 4 20 FS ...

Page 8: ...lue Press to select division value then press to confirm CALIBRATION Turn on the crane scale first then hold key and power on the indicator secondly it enters calibration mode Step Operation Display Explanation 1 Press to choose division value several times d Division value 1 2 5 10 20 50 optional press to confirm e g 2 2 Press to choose decimal point position several times P Decimal point optiona...

Page 9: ...weights loaded Press to choose the digit bit press to select the digit Until the input value is same as the loaded weight and the digit bit is at the highest bit press to confirm when the stable light is on 7 Calibration finished ERROR INDICATION 1 Err 1 AD value is too small when calibration 2 Err 2 Zero point is out of range when calibration 3 Err 3 Zero point is out of range when power on 4 Err...

Page 10: ...t be placed directly under sunshine and should be set in the plain space 2 The indicator can t be placed into the place where the dust pollution and vibration are serious 3 Load cell should connect with indicator reliably and the system should be well connected into ground The indicator must be protected from high electrical field and high magnetic field In order to protect the operator indicator ...

Page 11: ...ction may happen 8 The storage is not granted the free repair guarantee because it is easily exhausted products In order to prolong using life please charge the cell fully before using it If you don t use the indicator for a long time you must charge the cell every two month and for eight hours each charging time Moving or installation must be carefully taken and must avoid strong vibration impact...

Page 12: ...ation defect of the material During this period GRAM PRECISION SL covers the manpower and the spare parts necessary to repair the scale This guarantee does not cover the failures caused by an inappropriate use or overload The guarantee does not cover the freight cost transport necessary to repair the balance ...

Page 13: ...13 NOTES ...

Page 14: ...14 NOTES ...

Page 15: ...15 NOTES ...

Page 16: ...Gram Precision S L Travesía Industrial 11 08907 Hospitalet de Llobregat Barcelona Spain Tel 34 902 208 000 34 93 300 33 32 Fax 34 93 300 66 98 comercial gram es www gram group com 003 19072018 ...