JB0112-014-INS-LABRevB07
Clean Tubing:
1. Disconnect the tubing from the unit.
2. Tubing should be rinsed thoroughly after every use by running hot tap water through it followed
by a solution of one part vinegar to three parts hot water.
3. Rinse with hot tap water and air dry.
4. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.
WARNING:
The waste in collection bottle may cause respiratory infection which must be handled
according to medical waste disposing regulation. Operator must wear gloves and a
medical class mask before cleaning the collection bottle, tubing and handle the wasting
unit and/or accessories.
9. MAINTENANCE
Inspect suction tubing and collection bottle for leaks, cracks, etc. before each use.
DANGER:
Electric shock hazard
Do not attempt to open or remove the suction cabinet. If service is required, return
your suction unit to an authorized service center. Opening or tampering with the unit
will void the warranty.
Bacteria Filter Change:
1. The bacteria filter should be replaced every two months. If overflow occurs, change
the filter immediately.
2. Remove the filter by disconnecting it from the tubing connected to the pressure adjusting
valve and the tubing connected to the collection bottle lid assembly.
3. Replace with a new JOHN BUNN bacteria filter and connect to above two tubing.
Additional filters can be purchased from an authorized dealer.