When using electrical products, especially when children are present, basic safety precautions

should always be followed. Read all instructions before using.

Important information is highlighted by these terms:

DANGER:

Urgent safety information for hazards that will cause serious injury or death.

WARNING:

Important safety information for hazards that might cause serious injury.

CAUTION:

Information for preventing damage to the product.

NOTE:

Information to which you should pay special attention.

READ ALL USER’S MANUAL BEFORE USING.

SAVE USER’S MANUAL

DANGER

To reduce the risk of electrocution:

1. Do not use while bathing.

2. Do not place or store product where it can fall or be pulled into a tub or sink.

3. Do not place in or drop into water or other liquid.

4. Do not reach for a product that has fallen into water. Unplug immediately.

WARNING

To reduce the risk of burns, electrocution, fire or injury to persons:

1. Close supervision is necessary when this product is used by, on, or near

children or invalids.

2. Use this product only for its intended use as described in this guide.

3. Never operate this product if:

a. It has a damaged power cord or plug.

b. It is not working properly.

c. It has been dropped or damaged.

d. It has been dropped into water.

4. Return the product to an authorized JOHN BUNN medical equipment service center

for examination and repair.

5. Keep the power cord away from heated surfaces.

6. Never use while drowsy or asleep.

SPECIFIC WARNINGS

TO REDUCE THE RISK OF ELECTRICAL SHOCK OR INJURY TO PERSONS, DO NOT

DISASSEMBLE THIS UNIT. Disassembly or attempted repairs if accomplished incorrectly

can create electrical shock hazard. Only qualified technicians should perform servicing of

this suction unit.

1. SAFETY INFORMATION

Summary of Contents for JOHN BUNN JB0112-014

Page 1: ...hich he practices to use or order the use of this device NOTE Check ALL parts for shipping damage If shipping damage is noted DO NOT use Contact medical equipment provider for further instruction JB01...

Page 2: ...3 Electrical 3 4 Introduction 4 5 Illustration of Important Parts 5 6 Set Up 6 7 How to Operate Your JOHN BUNN Suction Unit 6 8 Cleaning Instructions 9 9 Maintenance 10 10 Troubleshooting 12 11 Speci...

Page 3: ...id 4 Do not reach for a product that has fallen into water Unplug immediately WARNING To reduce the risk of burns electrocution fire or injury to persons 1 Close supervision is necessary when this pro...

Page 4: ...operated appliances fail to operate properly Too many motor operated appliances on same circuit NOTE The motor of this compressor has a thermal overload protector If the motor overheats the overload...

Page 5: ...es New anti bacteria filter design helps reduce contamination High quality vacuum regulator and anti vibration vacuum gauge setting Oil less permanent lubrication Collection bottle with overflow prote...

Page 6: ...TANT PARTS MODEL 1 Vacuum Adjust Knob 2 Vacuum Gauge 3 Bacteria Filter 4 Connection Tubing 5 Cover 6 Patient Tubing 7 800cc Collection Bottle 8 Holder 9 Fuse 10 Power Switch 11 Power Cord 12 Base 13 R...

Page 7: ...secure and without leaks before using 7 HOW TO OPERATE YOUR JOHN BUNN SUCTION UNIT 1 Before use check the specification label on the side of the suction unit to ensure that the voltage and current in...

Page 8: ...e vacuum gauge while setting the desired level of vacuum To accurately read the gauge block the patient end of the hose or cap off the collection bottle and allow the gauge to reach a stable vacuum re...

Page 9: ...un the suction unit again NOTE 1 If the unit sustains a severe drop the validity of the gauge must be checked Inspect if the float shut off is still functioning properly when the liquid in the collect...

Page 10: ...ution and rinse with clean hot tap water Then wash in one part vinegar to three parts hot water solution Rinse with hot tap water and air dry b Wash with rubbing alcohol and air dry c Wash with a comm...

Page 11: ...d handle the wasting unit and or accessories 9 MAINTENANCE Inspect suction tubing and collection bottle for leaks cracks etc before each use DANGER Electric shock hazard Do not attempt to open or remo...

Page 12: ...broken fuse 2 Fuse specification Miniature Glass Fuse 5x20mm Fast Blow Quick Acting F2A 250V WARNING When this unit or accessories reach the end of their useful lives depends on using conditions resi...

Page 13: ...to check for these possible causes PROBLEM ACTION Unit does not turn on 1 Check power source and connection 2 Ensure that the wall outlet is live 3 Check power cord for damage Motor runs but no vacuum...

Page 14: ...4 70 watts Bacteria Filter Pore Size 0 45 microns Collection Bottle Capacity 800 ml cc Operating Temperature 32 104 F 0 40 C Operating Humidity 95 RH Operating Atmospheric Pressure 10 2 15 4 psi 70 10...

Page 15: ...FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED THIS IS THE EXCLUSIVE REMED...