Calibrate
Calibrate
Calibrate 2:
2:
2: Air
Air
Air and
and
and Fluid
Fluid
Fluid
Use the Calibrate 2 screen to adjust the calibration
value of the air and fluid. (Default value is 1000 mV.)
Calibrate the air and fluid:
• When the actual air or fluid pressure does not
match the target value on the Home screen and it
seems like it should.
.
For example, when air is applied to the system and
the system is in Standby mode, both target and
actual values should be 0.
• After replacing the electro/pneumatic regulator
(V2P) (415).
• After replacing the V2P cable.
Follow the steps in
Calibrate the Air and Fluid, page 62
.
Air:
Air:
Air: Current air pressure in millivolts (mV) for
calibration purposes.
Fluid:
Fluid:
Fluid: Current fluid pressure in mV for calibration
purposes.
Calibrate
Calibrate
Calibrate the
the
the Air
Air
Air and
and
and Fluid
Fluid
Fluid
Before calibrating the air and fluid, be sure that no air
pressure is present in the system.
1.
Perform the zero clear steps described within
Replace an Electro/pneumatic (V2P) Regulator,
page 110
.
2.
On the Calibrate 2 screen, select Air:
3.
On the Calibrate 2 screen, select Fluid:
Resume system use.
Calibrate
Calibrate
Calibrate 3:
3:
3: Bleed
Bleed
Bleed Resistor
Resistor
Resistor
Use the Calibrate 2 screen to adjust the
calibration value of the bleed resistor.
(Default value is 10000 Mohms.) Follow the
Calibrate the Bleed Resistor, page 62
.
Adjusting the value changes the kV display on the
Home screen.
Bleed
Bleed
Bleed Resistor:
Resistor:
Resistor: The resistor that is used to slowly
discharge the system, providing a path to ground
when the system is charged.
Note
All systems are shipped with the value set at
10,000 Mohm. A new bleed resistor displays
accurate system voltage.
Calibrate
Calibrate
Calibrate the
the
the Bleed
Bleed
Bleed Resistor
Resistor
Resistor
1.
With the bleed resistor (24) removed from the
system, measure the resistance across the bleed
resistor. See
Check the Bleed Resistor, page 73
.
2.
Use the keypad to enter the measured value.
Only values between 9,000 and 11,000 Mohm
are allowable.
3.
Reinstall the bleed resistor into the system.
4.
Reattach the wire from the fluid pressure
regulator (21) to the bleed resistor (24) and the
bleed resistor to the ground rod (18).
5.
Make sure the wire coming out of the top of the
bleed resistor is plugged back into the kV board
(428) inside the electronics control panel.
6.
After the reassembly is complete, check the
continuity between the fluid pressure regulator,
bleed resistor, and ground rod.
62
3A7370C