Low Level Fault/Warning
When approximately one lubrication cycle remains, the
blinking, yellow, low level warning light on the pump
panel, illuminates and Er:LL appears on the display as
shown in F
. 12. Fill the reservoir as soon as possible.
If the pump reservoir is not refilled with grease, the
pump will stop. A blinking, red, low level fault light and
Er:LL appears on the display as shown in F
Pumps with a remote run button will indicate the low
level warning and low level fault as shown on the Low
Level Fault/Warning Table.
Low Level Fault/Warning Table
Fuses
Graco Fuses
Pump Over-Current Protection
The pump has built-in over-current protection to protect
the motor. In cold weather installations, if the grease
freezes causing the motor to draw more current; the
motor will try to start the pump 90 times before shutting
down.
How Over-Current Protection Works
When the pump is unable to start the ON Time cycle, an
over-current condition is created:
a.
The pump immediately goes into the set OFF
Time mode.
b.
At the end of the OFF Time mode the pump
again attempts to run the ON Time cycle. One
of two results are possible at this point:
• the pump runs the ON Time cycle as
scheduled and normal cycle operation
begins.
• the pump is still not able to start (condition is
still in over-current) and the immediately goes
into the programmed OFF time again. This
will repeat for 90 times before shutting down
entirely.
F
IG
. 12
F
IG
. 13
Warning or Fault
Illumination
Low Level Warning
Yellow blinking
Low Level Fault
Red blinking
NOTICE
Fuses (user supplied) are required on all pump
models. To avoid equipment damage:
•
Never operate pump without a fuse installed.
•
A fuse of the correct voltage must be installed in
line with the power entry to the system.
Part No.
Description
Amps
25C985
12 VDC
10 Amps
25C986
24 VDC
7.5 Amps