4.6
Estimated Capacitance Measurement
Capacitance values can be determined in the resistance ranges
by estimated measurement. When doing so, proceed exactly as
if measuring resistance in accordance with chapter 4.5. Resis-
tance R
x
is to be replaced by the capacitance to be measured,
after prior discharge. When the capacitor is connected, the
pointer of the instrument swings to a maximum value and then
returns to its initial position (mechanical zero point). The point of
return of the pointer deflection serves as a measure for capaci-
tance. It is to be determined on scale 0 ... 5 V, A
. The mea-
sured value can be determined using the following comparative
scale and the factor for capacitance measurement correspond-
ing to the measuring range selected:
Before measurement is repeated, the capacitor must be
recharged!
Example:
Selected measuring range:
Ω
x 100
Return point of pointer:
3.3 on the upper scale 0 ... 5 V, A
Capacitance determined
via comparative scale:
50 μF
Multiplied by the factor for
capacitance measurement:
50 μF x 10 = 500 μF
Measuring Range
Factor for Capacitance
Measurement
Measuring Limits
Ω
x 1000
Ω
x 100
Ω
x 10
Ω
x
1
μF x
1
μF
x
10
μF x
100
μF x
1000
2 ...
200 μF
20 ...
2000 μF
200 ...
20000 μF
2 000 ...
200000 μF