Attention!
!
Only voltage-free objects may be measured. External
voltages falsify the measuring results and might also
damage the instrument!
➭
Read off the value displayed on the
Ω
scale and multiply by
the factor according to the adjusted measuring range.
If possible, the measuring range should be selected in such a
way as to obtain a reading in the range 5 ... 50
Ω
. The measuring
error, in relation to the actual resistance value, is smallest in the
middle of the deflection range. Read off the value displayed on
the
Ω
scale and multiply by the factor according to the adjusted
measuring range. During resistance measurements of a longer
duration, the full deflection 0
Ω
should always be checked if pos-
sible after moving the range switch from one resistance range to
another and, if necessary, it should be reset.
Note!
Contact resistances at the battery connecting terminals
may, particularly in low ohmic resistance ranges, cause
the setting of full deflection 0
Ω
to fluctuate. Conse-
quently, good contact should be ensured, for example
by removing and refitting the battery (see chapter 4.2).