GMC-I Gossen-Metrawatt GmbH
Balancing the Measured Current Final Value
Device settings:
OUTPUT ON; USET 2.00; ISET = I
nom
Output:
Shorted with shunt
– Measure current I
s
at shunt R
s
and balance the measured
Iout value which appears at the right-hand display with
R663a to:
Iout
= I
s
±
28 mA
where I
nom
= 12.5 A
= I
s
±
58 mA
where I
nom
= 25 A
= I
s
±
105 mA
where I
nom
= 50 A
= I
s
±
220 mA
where I
nom
= 75 A
= I
s
±
300 mA
where I
nom
= 100 A
= I
s
±
470 mA
where I
nom
= 150 A
Balancing the Voltage Setpoint Zero-Point
Device settings:
OUTPUT ON; USET 0.10; ISET 2.00
Output:
Idle
– Measure output voltage U
A
with the digital voltmeter and
balance with R554 to:
100 mV
±
5
mV
where
U
nom
= 40 V
100 mV
±
8
mV
where
U
nom
= 52 V
100 mV
±
10 mV
where U
nom
= 80 V
Balancing the Voltage Setpoint Final Value
Device settings:
OUTPUT ON; USET = U
nom
; ISET 2.00
U
A
= 40.00 V
±
30 mV
where U
nom
= 40 V
= 52.00 V ± 43 mV
where U
nom
= 52 V
= 80.00 V ± 60 mV
where U
nom
= 80 V
Balancing the Measured Voltage Zero-Point
Device settings:
OUTPUT ON; USET 0.00; ISET 0.00
Output:
Shorted
– Balance measured voltage value Uout, which appears at the
left-hand display, with R 657 to:
– 0.01
≤
Uout
≤
0.01
where U
nom
= 40 V
– 0.02
≤
Uout
≤
0.02
where U
nom
= 80 V or 52
V
Balancing the Measured Voltage Final Value
Device settings:
OUTPUT ON; USET = U
nom
; ISET 2.00
Output:
Idle
– Connect the digital voltmeter to the output and measure
voltage U
A
.
– Balance measured voltage value Uout, which appears at the
left-hand display, with R 651 to:
Uout =
U
A
±
20 mV
where U
nom
= 40 V
Uout =
U
A
±
23 mV
where U
nom
= 52 V
Uout =
U
A
±
40 mV
where U
nom
= 80 V
I m
o
nitor max.
I monitor 0
PCB FF