GMC-I Gossen-Metrawatt GmbH
33
rEP – Sequence Repetitions
Functions
•
Defines the number of
SEQUENCE
repetitions from the
start
address to the
stop
address.
•
The
SEQUENCE
is run from the selected start address to the
stop address. If repetitions have been programmed, the
pointer automatically jumps from the stop address back to the
start address, and the sequence is started over again.
•
The
rEP
value is stored to
SETUP
memory as a device setting
with the
<SAVE>
key.
Setting Parameters
cont
Continuous repetition until the
SEQUENCE
is stopped
manually (page 36 and page 36) or by means of
computer control (page 71)
1 ... 255
Number of possible repetitions
Settings
(see sequence settings and execution on page 35)
☞
Procedure
for selecting
rEP:
Settings.
!
The
rEP
display appears in the window for the SEq function,
along with a related numeric parameter [s].
Figure 4.7.3.5 a
Path to rEP Setting
☞
Select the desired number of repetitions with the <
↓
> key or
the <
↑
> key.
☞
After selecting the value, jump to the next or the previous
sequence function by pressing the
<FUNCTION>
key, or by
simultaneously pressing the
<FUNCTION>
key and the <
↓
> key
or the <
↑
> key.
!
The last selected and displayed value for the numeric
parameter (number of repetitions) is used as a default setting.
From the SEq function group
Select a value for
number of repetitions
FUNCTION
+