GOM-804 & GOM-805
使用手册
1
3
9
Example
TEMP:CONV:RES 10.00,maohm
Sets initial resistance value to 10.00MΩ.
TEMP:CONV:RES?
>10.0000E+6
Returns the initial resistance as 10.00MΩ.
TEMPerature:CONVersion:TEMPerature
Set
Query
Description
Sets or returns the initial temperature for the
temperature conversion function.
Syntax
Query Syntax
TEMPerature:CONVersion:TEMPerature <NRf>
TEMPerature:CONVersion:TEMPerature?
Parameter
<NRf>
-50.0~399.9 (Unit: º
C)
Return parameter
<NR2>
-50.0~399.9 (Unit: º
C)
Example
TEMP:CONV:TEMP 25.6
Sets the initial temperature to +25.6º
C.
TEMPerature:DATa
Query
Description
Returns the PT-100 sensor temperature measurement in
degrees Celsius.
Query Syntax
TEMPerature:DATa?
Return parameter
<NR3>
-50.0~399.9
Example
TEMP:DAT?
>0.250E+2
Returns the temperature as 25º
C.
TEMPerature:STATe
Set
Query
Description
Sets or returns the temperature function state.
Syntax
Query Syntax
TEMPerature:STATe {<NR1>|OFF|ON}
TEMPerature:STATe?
Parameter/
Return parameter
<NR1>
0:OFF
1:ON
OFF
Turn the temp function off.
ON
Turn the temp function on.
Summary of Contents for GOM-804
Page 1: ...直流微欧姆计 GOM 804 GOM 805 使用手册 固纬料号 ISO 9001 认证企业 ...
Page 21: ...GOM 804 GOM 805 使用手册 2 1 设置 倾斜站立 倾斜 将支脚向前扳 如下所示 直立 将支脚向后扳 如下所示 ...
Page 38: ...GOM 804 GOM 805 使用手册 38 Internal trigger source ...
Page 57: ...GOM 804 GOM 805 使用手册 57 GUARD SOURCE SENSE SENSE SOURCE shielding ...