GOM-804 & GOM-805
使用手册
1
33
SYSTem:LFRequency
Set
Query
Description
Sets or returns the frequency setting for the line filter.
Syntax
Query Syntax
SYSTem:LFRequency {AUTO | 50 | 60}
SYSTem:LFRequency?
Parameter/
Return parameter
AUTO
The frequency setting for the line
filter is automatically detected.
50
The frequency is 50Hz.
60
The frequency is 60Hz.
Example
SYST:LFR 60
Sets the line frequency to 60Hz.
SYST:LFR?
>60Hz
Returns the line frequency as 60Hz.
SYSTem:LOCal
Set
Query
Description
Enables local control (front panel control) and disables
remote control.
Syntax
SYSTem:LOCal
Parameter
<None>
SYSTem:MDELay:DATa
Set
Query
Description
Sets or returns the measurement delay time.
Syntax
Query Syntax
SYSTem:MDELay:DATa <NRf>
SYSTem:MDELay:DATa?
Parameter/
Return parameter
<NRf>
0.000~100.000
Unit:ms
For values under 1s, the unit
resolution is 1ms.
For values above 1s, the unit
resolution is 0.1s.