PSS-3203/2005 PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯
10
⎯
5. OPERATION METHOD
5-1. Output Voltage/Current Setting
At first, ensure that the window now is at the Output Voltage/
Current setting or the output value display.
or
--Output Voltage Setting:
Press [Vset/Iset] to set the cursor to the position for voltage input,
and modify the setting value with wheel knob. Right now, can use
[F/C] to switch over the input of 1V step or 10mV step.
Example:
If want to set the voltage as 12.34V, first using the [F/C] knob to
switch the cursor to 10mV step and adjust the value to 34, then
switch the cursor to 1V step and adjust the value to 12 to complete
the modification.
>>
>>
PS. If the output is on now, the output voltage value will be varied
with the setting of the knob.
--Output Current Setting:
After pressing [Vset/Iset] to set the cursor to the position for
current input, modify the setting value with wheel knob. Right now,
use [F/C] to switch over the input of 100mA step or 1mA step.
PSS-3203/2005 PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯
11
⎯
Example:
If want to set the current as 1.234A, first using the [F/C] knob to
switch the cursor to 1mA step and adjust the value to 34, then
switch the cursor to 100mA step and adjust the value to 12 to
complete the modification.
>>
>>
Note: 1. When the load current through output terminal exceeds
the setting value, the instrument is operated in the C.C.
mode, if not exceeds the setting value, the instrument is
operated in the C.V. mode.
2. When the maximum output voltage is larger than 36V, its
adjustable step is 20mV the minimum while the
maximum output current is larger 3A, its adjustable step
is 2mA the minimum.
5-2.Over Voltage /Current Protection Setting
--Over Voltage Protection Setting:
Set to OVP SET window by pressing [MENU], modify the setting
value with the wheel knob and press [ENTER]. The modification
can be done by using the [F/C] knob to switch over the input of
1V step or 10mV step.
--OVP Status Clear Up:
When the output voltage exceeds the setting voltage, the output of
the instrument will be off and get into OVP mode by displaying
“OVP Error. Press “LOCAL” to reset” on the panel. Just press
[LOCAL] to clear OVP status and back to previous status.