Good Will Instrument PSS-2005 User Manual Download Page 10

    PSS-3203/2005 PROGRAMMABLE POWER SUPPLY

USER MANUAL

  14  

6. MAINTENANCE

The following instructions are executed by qualified personnel only.

To avoid electrical shock, do no perform any servicing other than the

operating instructions unless you are qualified to do so.

6-1.Fuse Replacement

If the fuse blows, the display will not light and the power supply will

not operate. The fuse should not normally open unless a problem has

developed in the unit. Try to determine and correct the cause of the

blown fuse, then replace only with a fuse of the correct rating and

type (refer to 3-2 of page 5). The fuse is located on the rear panel

(see Fig.4-2).

WARNING. For continued fire protection. Replace fuse

only with 250V fuse of the specific type and rating, and

disconnect power cord before replacing fuse.

6-2. Line Voltage Conversion

The primary winding of the power transformer is tapped to permit

operation from 100, 120, 220, or 230VAC, 50/60 Hz line voltage.

Conversion from one line voltage to another is done by change

AC selects switch.

The rear panel identifies the line voltage to which the unit was

factory set. To convert to a different line voltage, perform the

following procedure:

WARNING

    PSS-3203/2005 PROGRAMMABLE POWER SUPPLY

USER MANUAL

  15  

(1)

 

Make sure the power cord is unplugged.

(2)

 

Ch ang e th e AC selects switch  to the d esired lin e voltage

   position.

(3)

 

A change in line voltage may also require a corresponding

change of fuse value.    Install the correct fuse value as listed on

rear panel.

6-3. ADJUSTMENT AND CALIBRATION

--Preparation

(1)

 

30 minutes warm up before calibration.

(2)

 

Ambient temperature:23

±

5°C, Humidity: Under PH80%.

--Output Calibration Steps:

[Step 1]

Press [LOCAL] for 4 to 5 seconds interruptedly to appear Password

input window (vary with different models: PSS-3203: 3203, PSS-

2005: 2005), input the password number with the wheel knob, and

use the F/C button to shift the input position at the same time, then

press [ENTER]. When the password number has been input by

mistake, the system will jump out of the present mode.

[Step 2] Voltage Calibration Step

After getting into Calibration selection window, select Voltage

calibration with the knob and press [ENTER].

Summary of Contents for PSS-2005

Page 1: ...12 13 13 13 6 MAINTENANCE 6 1 Fuse Replacement 6 2 Line Voltage Conversion 6 3 Adjustment and Calibration 6 4 Cleaning 14 14 14 15 18 7 THE SYSTEM DIAGRAM AND DESCRIPTION 7 1 Block Diagram 7 2 The Ope...

Page 2: ...loured Blue or Black The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red If in doubt consult the instructions provided with the eq...

Page 3: ...0 4 4 1995 Surge Immunity EN 61000 4 5 1995 Conducted Susceptibility EN 61000 4 6 1996 Voltage Dips Interrupts EN 61000 4 11 1994 Safety Low Voltage Directive 73 23 EEC amended by 93 68 EEC IEC EN 610...

Page 4: ...0 21V 1 Voltage 3mV 5mV rating current 3 0A Load Effect Current 3mA 5mA rating current 3 0A Voltage 3mV Source Effect Current 3mA Voltage 10mV Current 1mA 2mA rating current 3 0A Resolution OVP 10mV V...

Page 5: ...ng from the factory Upon receiving the instrument please unpack and inspect it to check if there is any damage caused during transportation If any sign of damage is found notify the bearer and or the...

Page 6: ...32 to 104 F To operate the instrument exceeding this specific temperature range may cause damage to the circuits of instrument Do not use the instrument in a place where strong magnetic or electric f...

Page 7: ...for value input or setting confirmation 8 F C The knob for switching over coarse and fine adjustment 9 MENU The category of function setting Output OVP OCP Contrast Buzzer Interface PS After switching...

Page 8: ...RAMMABLE POWER SUPPLY USER MANUAL 11 Example If want to set the current as 1 234A first using the F C knob to switch the cursor to 1mA step and adjust the value to 34 then switch the cursor to 100mA s...

Page 9: ...indow will be appeared then using the wheel knob to modify the value and press ENTER to complete the setting Note The system will detect the interface used at present automatically and switch the dete...

Page 10: ...tage to another is done by change AC selects switch The rear panel identifies the line voltage to which the unit was factory set To convert to a different line voltage perform the following procedure...