should be positioned directly in front of the object to be X-rayed and the imager placed directly behind
the object to be X-rayed. Imager should be placed as close to the object as possible. Distance between XR150
and imager is usually 24 to 72 inches (30 to 180 cm). During operation XR150 should be stabilized on a at
surface, a tripod, or a custom xture suitable for holding the 5 pound (2.2 Kg) XR150.
**OPERATING PRECAUTIONS:
The operator should always stand at least 10 feet (3m) behind the X-ray
unit and clear all personnel at least 10 feet (3m) behind the unit or at least 100 feet (30m) from the front of the unit
before pulsing. The exclusion zone (below) should be a controlled area free of all personnel while
X-ray pulses.
Figure 4: Exclusion Zone
24 to 72 inches (30 to 180
cm)
XR150
OBJECT
IMAGER
X-ray Beam
20’ (6m)
20’ (6m)
100’ (30m)
10’(3m)
10’(3m)
10’(3m)