INSTALACIÓN DEL EQUIPAMIENTO
Attention
The Gnatus dental set is dimensioned so that the stability of the set composed of a Dental
Chair, Delivery Unit, Dental Light and Water Unit remains stable in normal conditions of use.
If, by option of the customer, it is necessary to install other accessories in the dental set.
Gnatus informs you that the evaluation of the stability is the customer’s responsibility, and
suggests that in this case the base of the dental chair is fastened to the floor by means of
the holes already furnished in its base. This means that the added weight does not jeopardize
the stability of the set.
How to fix the dental chair base to the floor
Mark the floor, stick with 12mm drill, insert two S12 screw plastic plugs in the holes, place the chair
aligned to the holes and fix it using 3/8” x 90mm hex head steel screws and washers.
Base pierced for option of fastening
the dental chair to the floor: 2 Holes
located on the side of the bearings
“outside” covered by the finishing
cover of the motor.
Minina area for the installation
of a dental set should be 10m ².
Warning: Screws, washers and
bushings with the product.