7. Specifications
GME LinScan Revision 14, dated 2016-06-08
Page 43 of 46
No.
Message
Cause
User Action
down...
is too high.
range. Minimal cooling
time is achieved if de-
vice is kept running.
449
Applicator is cooling to config-
ured skin temperature. Please
wait..
Skin cooling temperature is
too high.
Please wait till temper-
ature reaches allowed
range.
450
Hand switch active error oc-
curred. Please release hand
switch!
Hand switch has already
been pressed before laser
was put in ready mode.
Release hand switch or
check if anything
blocks the hand switch.
451
Liquid flowrate low. Please refill
cooling liquid or straighten ca-
ble!
Not enough cooling liquid is
pumped through the device.
Two causes are possible:
-
There is not enough
cooling liquid in the
device
-
The applicator cable
is squeezed or
twisted
Fill up cooling liquid.
Check if cable to appli-
cator is bent heavily or
twisted. Straighten ca-
ble!
Restart device. If error
persists call service.
460
Creation of IO_READ_THREAD
failed
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
461
Creation of
IO_EMISSION_THREAD failed
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
462
Applicator error
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
463
bootloader in error state
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
464
erase flash failed
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
465
firmware upload failed
Hardware or software com-
ponent defective
Only for service
466
firmware crc error
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
467
start application failed
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
468
File not found
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
469
Queue overflow
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
470
general purpose input/output er-
ror
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
471
Unallowed temperature differ-
ence at applicator
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
472
Thermoelement Base Unit:
Open circuit
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
473
Thermoelement Base Unit:
Short circuit
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-