4. Installation
GME LinScan Revision 14, dated 2016-06-08
Page 13 of 46
4. Installation
4.1 Scope of delivery and Unpacking
The unit is supplied in a special carrying case, which in turn is put into an outer package for transpor-
tation by mail or parcel service. Upon delivery, please remove the carrying case and any accessories
from the outer packaging. Please open the carrying case carefully so that no parts can fall out.
Please check that all parts
–
according to the delivery note - are actually included and undamaged. A
complete delivery - ordered without any additional optional accessories - includes the following com-
ponents:
x
Base unit with an attached applicator
x
User manual (this document)
x
Quick Treatment Guide
x
Packaging Instruction
x
Power cord
x
Foot switch with connecting cable
x
One pair of laser protective goggles
x
Applicator holder
x
Refill kit
Upon damaged or missing parts, please immediately contact the manufacturer or distributor. Do not
use the device under any circumstances!
If all parts are contained place the base unit with the applicator on a stable surface. Follow the installa-
tion requirements specified in the following chapter.
Keep the outer packaging and packing materials. The carrying case and outer packaging may be re-
quired for the transport or shipping of the device. Do not ship or transport the device without the ap-
propriate box!
Caution!
Hazard:
The transport or shipping of the device without appropri-
ate packaging may damage the device and thus lead to a malfunc-
tion of the device.
Action:
Transport or ship the device only in transport cases and
packaging cleared by the manufacturer. If you do not have such
case or such packaging, please contact the manufacturer or your
dealer. He will provide you with suitable means of transport.
4.2 Installation Requirements
Before installing this equipment, users must make sure that the following requirements are fulfilled at
the intended treatment room. In addition, the user has to comply with all legal requirements, which are
valid at the location of use.
x
Required installation location: The device should be placed on a table. It must not be placed
directly against a wall, neither on the left nor right side, otherwise the ventilation is blocked.
On both sides, a distance of 50 cm to the next surface is required.
x
Separation from other electrical devices: The equipment must not be placed directly above,
below or next to other electrical devices. Excluded are devices or accessories that are specif-
ically provided by GME for such an assembly or have been released for this purpose by
GME.
x
Electrical connection: The appliance should be plugged into an outlet with its own fuse pro-
tection. It must not be connected to multi-outlet power strips or other distribution outlets. An
appropriate power cord with a protective earth is provided by the manufacturer.
!