Use only
TaiDoc
control solutions.
Check the expiry date on the control solution vial. Do not use if
expired.
Control solution, meter, and test strips should come to room
temperature (2°C-25°C) before testing.
Shake the vial, discard the first drop of control solution, and wipe off
the dispenser tip to ensure a good sample and an accurate result.
Use only for 3 months after first opening. Record the first opening
date on the control solution vial. Discard after 3 months.
Store the control solution tightly closed at temperatures 2°C to 30°C
(36°F to 86°F). Do not freeze.
PLEASE NOTE
The control solution range printed on the test strip vial is for
TaiDoc
control
solution only. It is used to test meter and test strip performance. It is not a
recommended range for your blood glucose level.
Important Control Solution Information