10. Troubleshooting Tips (Cont.)
IN THE CASE OF POWER OUTAGE
The lid will not open automatically if there is a power outage. It is possible to open manually as follows:
1. Ensure the rotor has stopped rotating.
• Listen carefully to ensure no rotation can be heard.
2. Insert a screw driver into each lid release hole to open the lid.
• Holes are located on the left and right sides of the unit.
• Carefully insert a screw driver into the lid release holes and push forward to release the lid.
10.3 Replacing Fuses
1. There are two fuses in the GCM-24: 250V, 5A time-delay type, size:
ф
5×20. There are two fuses of GCM-R-24: 250V,
10A time-delay type, size:
ф
5×20.
2. The fuse holder is located in the power inlet.
• Pull out the fuse holder tray from the power inlet and replace the fuses if necessary.
11. Rotor Instructions
CAUTION
• Read the instructions thoroughly to properly load and use the rotor.
• Do not exceed the allowable maximum speed of the rotor, tube and adapters etc. Ensure the allowable
maximum speed of the adapters is lower than the rotor’s maximum speed.
11.1 Rotor Structure
Knob
Sealing ring
Locking Nut
Cover
Shaft Hole