Precaution
s f
o
r u
se
Select a level and sturdy floor for installation. This precaution will prevent the unit from
tipping. Improper installation may result in water spillage or injury from the unit tipping
over.
Check the filter mentioned in this manual and clean it as necessary. A dusty filter may
cause temperature rise or failure.
Do not tilt the unit more than 45 degrees when moving the unit. All transportation should
be carefully.
u
When the freezer is running, the front part of the freezer may get hot. This is not a
malfunction: In order to prevent dew condensation around the case body, a heated anti-
condensation tube is installed in the cryogenic storage case.
u
Before putting the items in the freezer, please make sure the temperature inside the box has
reached the set temperature, and then put the items in batches. Add no more than 1/3 of the
container volume each time, in case the temperature rises too much.
u
The temperature display value shows the temperature of temperature sensors inside the
freezer. Although the displayed temperature may sometimes differ from the actual
temperature in the center of the freezer, it will get close to the real temperature gradually.
u
A through-hole is installed on the right or back of the freezer to allow test lines to exit the
test box. After drawing out the test lines, be sure to re-plug the test hole with insulating
material, otherwise the temperature inside the box will not be completely lowered, resulting
in condensation around the outside of the hole.
u
Clean the freezer with diluted neutral detergent. Do not use brushes, acid, gasoline, soap,
polish or hot water to clean the frozen storage container, as these materials may damage
the painted surfaces and plastic rubber components.
u
After a period of operation, a layer of frost will form on the inner wall and the inner door. If
the frost layer is too thick, it will affect the insulation effect of the freezer and increase the
power consumption. Therefore, after a period of time, when the frost layer reaches about
5mm, it should be defrosted with the defrost shovel attached.
u
Please remove the frozen items in the freezer and put them in a suitable storage
environment before defrost, so as to avoid the damage of goods for temperature rise.
u
There are many cooling coils on the back and side of the inner wall. Don't use sharp objects
such as knives, ice chisel, or screwdrivers to remove the frost on the inner wall. Be careful
not to scratch the inner wall when defrosting, otherwise it will cause a failure of the freezer .
u
Cut off the power when the refrigerated storage box will not be used for a long time.
5.
Components and Overview
Dispaly control panel
Cabinet
Adjustable foot
Movable casters
Door handle
Door lock
Door handle
1. Due to the improvement of the product and the different models, the actual product may
be different from the diagram. The above figure is only an example of the identification function.
2. Structure and composition: Cabinet, door body, refrigeration system, control system and
accessories. Medical refrigerators and freezers are suitable for hospitals, epidemic prevention
stations, blood banks, research institutes, universities, genetic engineering, companies, etc.
3. The storage temperature inside the box can be adjusted through the temperature button
on the control panel, which is convenient to use and reliable in performance.
Door