Glacier GL-F-16M Operation Manual Download Page 12

9

10

Power lock 
and Back-up 
Battery lock

(do not open it till power failure)

Open to clean 
the Condenser 
filter net

Inner door

Cabinet inner 
seal

Test port

Shelves

Shelves frame

Outer door 
seal

Inner door 
seal

Just to switch on the power supply for 

medical freezer

 do not immediately add items, make empty operation after a 

period of time (about 24 hours), then refrigerated goods into storage box.

To add items, if the item is too much water or excessive drying will influence the change of the humidity, the best 

goods sealed; the same 

medical freezer

 working environment humidity size will also affect the change of the humidity 

of the door and the door body ,especially too much not close well.

For each store items can not exceed 1/3 box capacity, achieve the actual temperature and normal operation after 24 

hours, then add another 1/3 box storage capacity.

To add items please do not block the outlet and inlet.

       

The goods can not be placed directly on the bottom of the 

medical freezer

 should be placed on the bottom shelf, 

otherwise the effect of 

medical freezer

 cooling effect.

The cooling process as far as possible not to open the door, otherwise it will cause temperature rise.

Due refrigeration inertia, the product reaches a set value when stopping the display temperature and the set temperature 

may have a certain difference

, this is the normal phenomenon.

Note: children are not allowed to play games in a 

medical freezer

.

Caution

For personal safety, please cut the power supply before any maintenance!

Don't inhale medications or aerosols around the device while maintaining a 

medical freezer

, or it

 

will endanger

your health.

7

.

Defrost

 ●

Unused

 ● 

Maintenance

     After the product has been working for a period of time, its internal surface will be frosted, which will affect the 
cooling effect and increase the power consumption. It is recommended that defrosting should be performed when 
the frost layer is too thick.
     Cut off the power, open the door, and transfer the items to a cool place.
     Use a defrost scoop to gently remove the surface area cream (Also allow the temperature inside the tank to 
naturally rise, melt the frost)
      In order to speed up the defrosting process, flat bottom containers with less than 50°C hot water can be placed in 
a medical freezer.
      Then use a dry cloth to wipe off any remaining ice water. Return items to the medical refrigerator.
      Power on, power on cooling.

7.

1 D

efrost

Warning: Do not use sharp metal appliances during defrosting to 
avoid damage to the medical freezer.

7.2

 U

nuse

d a

n

d M

aintenance

      If the product needs to be stopped for a long time, the power should be cut off, cleaned 
according to following methods, and sealed after opening and drying. 
      Before using it again, apply a dry cloth to dry the water around the door.

Summary of Contents for GL-F-16M

Page 1: ...www aegisfridge com OPERATION MANUAL Applicable product model GL F 16M ...

Page 2: ... 4 4 Standing 4 5 Power on for the First Time 5 Components and Overview 6 Display Panel Function Description 7 Defrost Unused Maintenance 7 1 Defrost 7 2 Unused and Maintenance 7 3 Disposal 8 After sale Service 9 Specifications 10 Performance 10 1 Main Technical Parameter 10 2 Packing List 10 3 Electrical Diagram 1 2 4 4 4 4 5 5 6 10 10 12 13 15 15 15 15 5 7 10 11 ...

Page 3: ...carefully the cautions and safety issues In order to make the device work properly please follow the correct installation method Medical freezer ultra low temperature freezer combined refrigerator and freezer can only be operated by trained and authorized personnel Maintenance of the equipment can only be carried out by agents authorized by Glacier If the operator encounters a situation not mentio...

Page 4: ...onnel may cause electric shock or fire Be sure to install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipping over Carefully with the power cord to avoid short circuit or open circuit When removing the plug from the power supply outlet grip the power supply plug not the cord Pulling the cor...

Page 5: ...or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit Use a power supply outlet with ground earth to prevent electric shock If the power supply outlet is not grounded it will be necessary to install a ground by qualified engineers The medical freezeris not available to store non living things flowers or other critical articles which is not suitable for low t...

Page 6: ... strong vibration and no corrosive gases around Without the presence of a lot of dust No shaking or vibrating of the freezer Altitude of the place where the freezer is located 2000m Input voltage 220 10 V No direct sunlight or any other cooling or heating source no electromagnetic interference or the freezer will not run properly 4 2 Installation site This unit is a precision machine When select a...

Page 7: ...d transfer to other safe locations 5 Prior to put the articles inside the freezer should check if the temperature set range of this device is matching the requirement of the articles 6 Due to the Inertia of refrigeration there is a little difference between the actual temperature displayed on the controller and the set temperature This is a normal phenomenon 7 Medical freezers are storage devices ...

Page 8: ...ample of the identification function 2 Structure and composition Cabinet door body refrigeration system control system and accessories Medical refrigerators and freezers are suitable for hospitals epidemic prevention stations blood banks research institutes universities genetic engineering companies etc 3 The storage temperature inside the box can be adjusted through the temperature button on the ...

Page 9: ...e set temperature increase the set temperature value When setting the parameter value press and hold the up button for a long time and the parameter will increase rapidly Under normal conditions press and hold the key for 3 seconds to import the U disk for 12 months g is the button for set mute when there is no alarm state and the key is not unlocked press displaying ambient temperature for 5 seco...

Page 10: ...eter settings Unlock In the normal running state press the key and for 3 seconds at the same time the upper room digital tube will display the parameter code PS1 the lower room digital tube will display the parameter code 0000 press the and input the password 0005 enter the user menu password enter 0099 to restore the key lock password to the default 0005 Unlock at this time Press the key for 3 se...

Page 11: ...wer chamber low temperature alarm High ambient temperature alarm U disk generated file error Upper chamber temperature sensor alarm Low battery alarm door PF Door ajar alarm Power failure alarm EE Communication failure alarm Note When adding USB or printer ER recorder not connected and LoF recorder not activated abnormal alarm are valid 5 L01 0 0 10 0 5 0 Low temperature alarm setting value is set...

Page 12: ...re not allowed to play games in a medical freezer Caution For personal safety please cut the power supply before any maintenance Don t inhale medications or aerosols around the device while maintaining a medical freezer or it will endanger your health 7 Defrost Unused Maintenance After the product has been working for a period of time its internal surface will be frosted which will affect the cool...

Page 13: ...r with a soft cloth Once this product is activated it is best to use it continuously Note Do not spray water directly on the cabinet so as to avoid degradation of electrical components and metal parts rust Do not use hot water and corrosive cleaners or organic solvents to clean the cabinet 7 3 Disposal WARNING If the equipment is left unused in an unsupervised area for extended periods of time mak...

Page 14: ...nnected Is the power supply voltage too low or too high Whether the door is closed or the door is open too often Do you have too many items at one time Is the ambient temperature too high Temperature setting is correct or not Is the temperature inside the box too low Is the box placed on a flat floor Does the cabinet hit the wall During the rainy and humid season the window may condense which is n...

Page 15: ...m filling Fully enclosed Medical freezer Ultra low temperature freezer GL F 16M PCM color plate Epoxy polyester powder spray coated steel plate Wire tube condenser Wire Tube Evaporator Microcomputer control NTC High low temperature alarm door ajar alarm power failure alarm low battery alarm sensor error alarm Thermostat Temperature sensor ...

Page 16: ...lace it with the same roll paper The installation steps are as follows 1 Press the cylinder button on the printer and open the cover of the printer 2 put print paper into the printer box and pull the paper roll end of the printer cover after the note slightly exposed on the box cover port Annex1 Rechargeable Battery Maintenance Installation Replacement and Disposal Long term power failure or in th...

Page 17: ...nical Parameter Model Ambient tempera ture Refriger ant Rated voltage V Rated frequenc y Hz Storage temperatu re Net capacity L Rated current A Net Weight Kg Exterior dimension W D H mm Clim ate type Name Operation manual 1 10 3 Electrical Diagram Quantity Defrost shovel 1 10 2 Packing List GL F 16M 16 32 N R600a 65g 2 220 50 Up 10 25 450 1 55 144 810 735 1960 Down 10 25 ...

Reviews: