10 PC SETUP & MONITOR
The PC software is used to enter the initial machine data, X1, X2 etc. but can also be used
to monitor the condition of the machine, store the calibration settings and even adjust the
calibration settings as required.
10.1 PC Connection
The machine data etc. is entered
and sent to the Series 1-L Control
unit as described in the Initial
set-up section on page 13.
10.2 PC Indicator Monitoring
Click on the Calibrate heading
on the command line and the
Indicator screen, shown below,
appears.
The Calibrate box is inactive
(“Greyed Out”). The current
values of the Series 1-L Angle
& Pressure are shown, with
numbers representing the
current Load Moment and the
Capacity (Maximum Moment) of
the system. A Bar Column on the
left mimics the Green, Amber
& Red LEDs on the Series 1-L
Display unit. Increasing the load
will be shown as an increase in Load Moment. Increasing the boom angle (decreasing the
radius) will be shown as a decrease in Load Moment.