2.

 

General safety instructions

 

If the oven/incubator is not installed, operated, cleaned, adjusted or set up correctly, there is a risk of malfunction 
which  could  lead  to  personal  injuries  and  material  damages  to  the  instrument  and  samples.  Therefore,  the 
oven/incubator must only be installed operated, cleaned, adjusted and set up by qualified personnel.  
 

                    DANGER

 

 
 
 
 

 

 

Risk of electric shock and death 

 

Do not get the instrument wet during installation, commissioning or maintenance. 

 

Do not connect the instrument to power if the rear panel is dented or damaged. 

       

 

Before opening the rear panel, remove the power plug from the power supply. 

 

If the power cable or the rear panel of the instrument are damaged, stop using them 
immediately, remove the plug from the power supply and contact your dealer for the 
necessary repairs.  

 

All work on the electrical components of the instrument must be carried out by qualified 
personnel only. 

 

Danger of explosion 

 

Install the instrument only where there is no risk of explosion. 

 

Do not keep air/solvent mixtures or explosive dusts nearby. 

 

Never  introduce  materials  into  the  instrument  that  are  explosive  or  flammable  at  the 

selected operating temperature.

 

 

Never introduce materials containing flammable or explosive solvents into the instrument.

 

 

Never  introduce  into  the  instrument  materials  that  by  sublimation  or  pyrolysis  create 

flammable materials at the selected operating temperature. 

 

Danger of poisoning and death 

 

Never introduce materials into the instrument whose disintegration could create poisonous 
gases at the selected operating temperatures. 

 
 

       WARNING            

 

 

 

Fire hazard 

 

The ovens/incubators must not be used if the inspection of thermostat with safety class 2 
has failed. 

 

If the safety thermostat test fails, immediately stop using the oven/incubator, remove the 
plug from the power supply and contact your dealer for the necessary repairs. 

 

Always place the instrument on a working surface that is resistant to a temperature of 100 
°C. 

 

Do not put anything under the instrument (paper, plastic film, etc.). 

 

Always connect the instrument only to a power supply with a fuse of at least 10A. Follow the 
recommendations of your local power supply company. 

Summary of Contents for ICF-120 Plus

Page 1: ...OVENS and INCUBATORS User manual...

Page 2: ...troller Model Description Temperature range ICN 16 Plus Natural convection incubator 16L maximum volume Room temperature from 5 C to 70 C ICN 35 Plus Natural convection incubator 35L maximum volume Ro...

Page 3: ...gns 5 Symbols of obligation 5 2 GENERAL SAFETY INSTRUCTIONS 6 3 CE MARKING DATA 8 4 CONTENT OF PACKAGE 9 5 TRANSPORTATION 9 Instructions for safe transportation 9 Transportation of an already used ove...

Page 4: ...Cycle functions 15 Repetition of a selected program 15 Delay starts function 16 12 INTRODUCTION OF SAMPLES INTO OVEN INCUBATOR 16 Limit temperature for sample protection 16 Restart mode after absence...

Page 5: ...ions or incorrect results due to failure to comply with them Read entirely and carefully this manual and be sure to familiarize with the tool before starting to work with it This manual must be kept n...

Page 6: ...surfaces Danger of damage to health caused by toxic substances Danger of injury from overturning of objects Risk of injury from lifting heavy objects Danger of environmental damage Danger of corrosio...

Page 7: ...y where there is no risk of explosion Do not keep air solvent mixtures or explosive dusts nearby Never introduce materials into the instrument that are explosive or flammable at the selected operating...

Page 8: ...f each other Always fix 2 stacked ovens with the fixing plates provided Risk of injury Risk of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people Th...

Page 9: ...ors EN rev 0 31 05 2021 3 CE marking data Argolab instruments are manufactured in compliance with Directive 2006 42 EC and the relevant Community Directives applicable at the time of placing on the ma...

Page 10: ...rtation ATTENTION Risk of injury of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people The instrument must be transported in its original packaging...

Page 11: ...rmitted storage temperature is 10 C to 60 C The maximum permitted storage humidity is 85 RH without condensation 7 First installation Getting started The oven should be installed in following conditio...

Page 12: ...11 UM Ovens and Incubators EN rev 0 31 05 2021 8 Instrument parts Display and commands Display Adjustment keys ON OFF Key Opening closing handle...

Page 13: ...mperature variation at 150 C 0 3 C 0 3 C 0 4 C 0 5 C Heating time at 150 C 20 min 24 min 30 min 50 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 3 1 Po...

Page 14: ...at 37 C 40 min 45 min 55 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 Power supply power 230 V 600 W 230 V 700 W 230 V 1500 W Internal dimensions L A P 520 x 530 x 4...

Page 15: ...using the buttons in H High M Medium L Low otherwise go to STEP 2 STEP 2 Use the and SHIFT keys to set the temperature value of the second working STEP Briefly press the SET PROG button to confirm the...

Page 16: ...e this operation has been carried out the word Lk lock appears in the top right hand corner of the display in place of the word TIME next to the digits 0000 password Here below you can find the passwo...

Page 17: ...mit for sample protection The instrument foresees the possibility of limiting the maximum working temperature to protect the samples from an incorrect temperature setting of the heating cycle Follow t...

Page 18: ...ormal and therefore reducing the AL value excessively would risk causing the instrument to go into alarm frequently and unnecessarily Follow the instructions in paragraph 10 and use the keys to set th...

Page 19: ...corrosive DANGER Risk of electric shock and death DO NOT spray water or cleaning agents onto the interior and exterior surfaces Switch off the main switch and pull out the mains plug before cleaning D...

Page 20: ...does not apply to any product or parts thereof that have been damaged due to incorrect installation improper connection misuse accident or abnormal operating conditions No liability is accepted for d...

Reviews: