Giorgio Bormac ICF-120 Plus User Manual Download Page 14

 
 

13 

 

UM Ovens and Incubators EN rev. 0 31.05.2021 

Natural convection 
incubators 

ICN-16 Plus 

ICN-35 Plus 

ICN-55 Plus 

ICN-120 Plus 

ICN-200 Plus 

Volume  

16 l 

35 l 

55 l 

120 l 

200 l 

Max temperature/ Resolution 

+70/0,1°C 

+70/0,1°C 

+70/0,1°C 

+70/0,1°C 

+70/0,1°C 

Homogeneity at 37°C 

± 0,4 °C 

± 0,4 °C 

± 0,5 °C 

± 0,5 °C 

± 0,5 °C 

Temperature variation at 37°C 

± 0,3°C 

± 0,3°C 

± 0,3°C 

± 0,3°C 

± 0,3°C 

Heating time at 37°C 

18 min. 

22 min. 

25 min. 

30 min. 

35 min. 

Timer 

99:59 hh:min e 

 

99:59 hh:min e 

 

99:59 hh:min e 

 

99:59 hh:min e 

 

99:59 hh:min e 

 

Safety class 

Power supply/power 

230 V / 85 W 

230 V / 125 W 

230 V / 250 W 

230 V / 350 W 

230 V / 600 W 

Internal dimensions (L*A*P) 

270 x 230 x 255 mm  360 x 300 x 320 mm 

400 x 360 x 385 mm 

520 x 460 x 500 mm 

610 x 600 x 575 mm 

Shelves number 
(standard/max) 

2/3 

2/6 

2/5 

2/7 

2/9 

Distance between shelves 

25 mm 

30 mm 

50 mm 

50 mm 

50 mm 

Max loading of shelves 

5 Kg 

7,5 Kg 

10 Kg 

10 Kg 

10 Kg 

External dimensions (L*A*P) 

530 x 370 x 400 mm  620 x 440 x 460 mm 

660 x 500 x 545 mm 

780 x 610 x 645 mm 

875 x 755 x 710 mm 

Weight 

23 Kg 

33 Kg 

42 Kg 

61 Kg 

77 Kg 

 

Forced air incubators

 

ICF-120 Plus 

ICF-200 Plus 

ICF-400 Plus 

Volume  

120 l 

200 l 

400 l 

Max temperature/ Resolution 

+80/0,1°C 

+80/0,1°C 

+80/0,1°C 

Homogeneity at 37°C 

± 0,4 °C 

± 0,4 °C 

± 0,5 °C 

Temperature variation at 37°C 

± 0,1°C 

± 0,2°C 

± 0,3°C 

Heating time at 37°C 

40 min. 

45 min. 

55 min. 

Timer 

99:59 hh:min e 

 

99:59 hh:min e 

 

99:59 hh:min e 

 

Safety class 

3.1 

3.1 

3.1 

Power supply/power 

230 V / 600 W 

230 V / 700 W 

230 V / 1500 W 

Internal dimensions (L*A*P) 

520 x 530 x 435 mm 

645 x 650 x 495 mm 

1000 x 800 x 500 mm 

Shelves number (standard/max) 

2/7 

2/9 

2/10 

Distance between shelves 

50 mm 

50 mm 

50 mm 

Max loading of shelves 

20 Kg 

20 Kg 

20 Kg 

External dimensions (L*A*P) 

810 x 750 x 690 mm 

945 x 870 x 755 mm 

1285 x 1060 x 750 mm 

Weight 

74 Kg 

103 Kg 

160 Kg 

 

10.

 

Operation

 

 

Switching on the instrument

 

Connect the power cable to a grounded socket. Switch on the instrument using the ON/OFF button. The button and 
the display light up. 
The display shows the initialisation sequence and then the instrument is ready for use. 

NOTE:

 when the instrument is switched on, it emits an intermittent acoustic signal; the visual alarm icon   and the 

word 

"end" 

appear on the display, indicating that a heating cycle had been completed before switching it off. The 

acoustic signal can be muted by pressing any key, and the icon 

 appears on the display. 

 

 

Programming

 

Each Argolab oven/incubator can manage up to 7 programmes, each consisting of 10 working steps in which the 
temperature, timer and ventilation speed (where applicable) can be set. 
In addition to the above-mentioned programmes, there is "PROG 0", which can be used to set a simple operating 
cycle with a single working step, consisting of the following parameters: temperature, timer and ventilation speed 
(where applicable). 
 
 

Summary of Contents for ICF-120 Plus

Page 1: ...OVENS and INCUBATORS User manual...

Page 2: ...troller Model Description Temperature range ICN 16 Plus Natural convection incubator 16L maximum volume Room temperature from 5 C to 70 C ICN 35 Plus Natural convection incubator 35L maximum volume Ro...

Page 3: ...gns 5 Symbols of obligation 5 2 GENERAL SAFETY INSTRUCTIONS 6 3 CE MARKING DATA 8 4 CONTENT OF PACKAGE 9 5 TRANSPORTATION 9 Instructions for safe transportation 9 Transportation of an already used ove...

Page 4: ...Cycle functions 15 Repetition of a selected program 15 Delay starts function 16 12 INTRODUCTION OF SAMPLES INTO OVEN INCUBATOR 16 Limit temperature for sample protection 16 Restart mode after absence...

Page 5: ...ions or incorrect results due to failure to comply with them Read entirely and carefully this manual and be sure to familiarize with the tool before starting to work with it This manual must be kept n...

Page 6: ...surfaces Danger of damage to health caused by toxic substances Danger of injury from overturning of objects Risk of injury from lifting heavy objects Danger of environmental damage Danger of corrosio...

Page 7: ...y where there is no risk of explosion Do not keep air solvent mixtures or explosive dusts nearby Never introduce materials into the instrument that are explosive or flammable at the selected operating...

Page 8: ...f each other Always fix 2 stacked ovens with the fixing plates provided Risk of injury Risk of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people Th...

Page 9: ...ors EN rev 0 31 05 2021 3 CE marking data Argolab instruments are manufactured in compliance with Directive 2006 42 EC and the relevant Community Directives applicable at the time of placing on the ma...

Page 10: ...rtation ATTENTION Risk of injury of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people The instrument must be transported in its original packaging...

Page 11: ...rmitted storage temperature is 10 C to 60 C The maximum permitted storage humidity is 85 RH without condensation 7 First installation Getting started The oven should be installed in following conditio...

Page 12: ...11 UM Ovens and Incubators EN rev 0 31 05 2021 8 Instrument parts Display and commands Display Adjustment keys ON OFF Key Opening closing handle...

Page 13: ...mperature variation at 150 C 0 3 C 0 3 C 0 4 C 0 5 C Heating time at 150 C 20 min 24 min 30 min 50 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 3 1 Po...

Page 14: ...at 37 C 40 min 45 min 55 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 Power supply power 230 V 600 W 230 V 700 W 230 V 1500 W Internal dimensions L A P 520 x 530 x 4...

Page 15: ...using the buttons in H High M Medium L Low otherwise go to STEP 2 STEP 2 Use the and SHIFT keys to set the temperature value of the second working STEP Briefly press the SET PROG button to confirm the...

Page 16: ...e this operation has been carried out the word Lk lock appears in the top right hand corner of the display in place of the word TIME next to the digits 0000 password Here below you can find the passwo...

Page 17: ...mit for sample protection The instrument foresees the possibility of limiting the maximum working temperature to protect the samples from an incorrect temperature setting of the heating cycle Follow t...

Page 18: ...ormal and therefore reducing the AL value excessively would risk causing the instrument to go into alarm frequently and unnecessarily Follow the instructions in paragraph 10 and use the keys to set th...

Page 19: ...corrosive DANGER Risk of electric shock and death DO NOT spray water or cleaning agents onto the interior and exterior surfaces Switch off the main switch and pull out the mains plug before cleaning D...

Page 20: ...does not apply to any product or parts thereof that have been damaged due to incorrect installation improper connection misuse accident or abnormal operating conditions No liability is accepted for d...