Restart mode after absence of power energy

 

The mode in which the instrument starts working again after a power failure can be set: 

VALUE Po

 

DESCRIPTION

 

When the power supply returns, the instrument does not automatically resume the heating cycle 
but must be restarted manually. 

When  the  power  supply  returns,  the  instrument  automatically  resumes  operation  from  the 
beginning of the interrupted heating cycle. 

When the power supply returns, the instrument automatically resumes operation from the precise 
point in the heating cycle at which it was interrupted. 

Follow the instructions in paragraph 9 and use the keys 

 to set the password 0003. You can move quickly 

between  the  digits  using  the  SHIFT  key 

  .  Confirm  the  value  by  pressing  SET/PROG  button 

  again.  The 

parameter  "

tm

"  (max.  temperature)  appears  in  the  top  right-hand  corner  of  the  display,  move  on  to  the  next 

parameter "

Po

" (Power) by pressing SET/PROG button 

 again. Set the desired value (0, 1, 2) by pressing the keys 

. Confirm the value by pressing SET/PROG button 

 again. 

 

 

Temperature limit for over-temperature alarm

 

It is possible for the user to set the temperature value beyond which the instrument goes into overtemperature 
alarm. 

NOTE:

 although it can be modified by the operator, this value is already set at the factory and is specifically 

calibrated to the type of instrument in question, natural/forced oven or natural/forced incubator. 
It  is  therefore  advisable  not  to  modify  this  value  unless  strictly  necessary,  as  temperature  fluctuations 
above  or  below  the  set  value,  especially  in  natural  convection  models,  are  completely  normal  and  therefore 
reducing the AL value excessively would risk causing the instrument to go into alarm frequently and unnecessarily. 

Follow the instructions in paragraph 10 and use the keys 

 to set the password 0003. You can move quickly 

between  the  digits  using  the  SHIFT  key 

.  Confirm  the  value  by  pressing  SET/PROG  button 

  again.  The 

parameter  "

tm

"  (max  temperature)  will  appear  on  the  display  in  the  top  right-hand  corner,  briefly  press  the 

SET/PROG key 

 to move on to the next parameters. When you reach the 

AL

 (alarm) parameter, set the minimum 

temperature value above which you want the instrument to go into an overtemperature alarm by pressing the keys 

. You can move quickly between the digits using the SHIFT key 

Confirm the value by pressing SET/PROG button 

again. 

 

13.

 

Temperature offset on single point, on entire range,

 

on room temperature sensor 

 

The instrument allows the user to set the offset values, i.e., the calibration values, on a temperature point, on the 
entire temperature range and on the room temperature range. 

NOTE:

  although  they  can  be  modified  by  the  operator,  these  values  are  already  set  by  the  factory  and 

perfectly calibrated with certified measuring instruments and Accredia references. It is therefore advisable 
not to modify these values unless strictly necessary, for example, if a certified digital thermometer reveals 
inconsistencies between the temperature readings of the instrument and those taken by the thermometer itself. 

Follow the instructions in paragraph 10 and use the keys 

 to set the password 0003. 

You can move quickly between the digits using the SHIFT key 

. Confirm the value by pressing SET/PROG button 

 again. The parameter "

tm

" (max. temperature) will appear on the display in the top right-hand corner, briefly 

press SET/PROG button 

 to move on to the next parameters until the desired ones are reached. 

 
 

Summary of Contents for ICF-120 Plus

Page 1: ...OVENS and INCUBATORS User manual...

Page 2: ...troller Model Description Temperature range ICN 16 Plus Natural convection incubator 16L maximum volume Room temperature from 5 C to 70 C ICN 35 Plus Natural convection incubator 35L maximum volume Ro...

Page 3: ...gns 5 Symbols of obligation 5 2 GENERAL SAFETY INSTRUCTIONS 6 3 CE MARKING DATA 8 4 CONTENT OF PACKAGE 9 5 TRANSPORTATION 9 Instructions for safe transportation 9 Transportation of an already used ove...

Page 4: ...Cycle functions 15 Repetition of a selected program 15 Delay starts function 16 12 INTRODUCTION OF SAMPLES INTO OVEN INCUBATOR 16 Limit temperature for sample protection 16 Restart mode after absence...

Page 5: ...ions or incorrect results due to failure to comply with them Read entirely and carefully this manual and be sure to familiarize with the tool before starting to work with it This manual must be kept n...

Page 6: ...surfaces Danger of damage to health caused by toxic substances Danger of injury from overturning of objects Risk of injury from lifting heavy objects Danger of environmental damage Danger of corrosio...

Page 7: ...y where there is no risk of explosion Do not keep air solvent mixtures or explosive dusts nearby Never introduce materials into the instrument that are explosive or flammable at the selected operating...

Page 8: ...f each other Always fix 2 stacked ovens with the fixing plates provided Risk of injury Risk of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people Th...

Page 9: ...ors EN rev 0 31 05 2021 3 CE marking data Argolab instruments are manufactured in compliance with Directive 2006 42 EC and the relevant Community Directives applicable at the time of placing on the ma...

Page 10: ...rtation ATTENTION Risk of injury of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people The instrument must be transported in its original packaging...

Page 11: ...rmitted storage temperature is 10 C to 60 C The maximum permitted storage humidity is 85 RH without condensation 7 First installation Getting started The oven should be installed in following conditio...

Page 12: ...11 UM Ovens and Incubators EN rev 0 31 05 2021 8 Instrument parts Display and commands Display Adjustment keys ON OFF Key Opening closing handle...

Page 13: ...mperature variation at 150 C 0 3 C 0 3 C 0 4 C 0 5 C Heating time at 150 C 20 min 24 min 30 min 50 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 3 1 Po...

Page 14: ...at 37 C 40 min 45 min 55 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 Power supply power 230 V 600 W 230 V 700 W 230 V 1500 W Internal dimensions L A P 520 x 530 x 4...

Page 15: ...using the buttons in H High M Medium L Low otherwise go to STEP 2 STEP 2 Use the and SHIFT keys to set the temperature value of the second working STEP Briefly press the SET PROG button to confirm the...

Page 16: ...e this operation has been carried out the word Lk lock appears in the top right hand corner of the display in place of the word TIME next to the digits 0000 password Here below you can find the passwo...

Page 17: ...mit for sample protection The instrument foresees the possibility of limiting the maximum working temperature to protect the samples from an incorrect temperature setting of the heating cycle Follow t...

Page 18: ...ormal and therefore reducing the AL value excessively would risk causing the instrument to go into alarm frequently and unnecessarily Follow the instructions in paragraph 10 and use the keys to set th...

Page 19: ...corrosive DANGER Risk of electric shock and death DO NOT spray water or cleaning agents onto the interior and exterior surfaces Switch off the main switch and pull out the mains plug before cleaning D...

Page 20: ...does not apply to any product or parts thereof that have been damaged due to incorrect installation improper connection misuse accident or abnormal operating conditions No liability is accepted for d...

Reviews: