14
3. Technical Specifications
A. Display mode: LCD Display
B. Power supply requirement:
1.5V (AAA size) alkaline battery × 2
Supply voltage: 3VDC
Operating current: <=15mA (backlight off)
C. SpO
2
Parameter Specifications
Measuring range: 35%~99%
Accuracy: ±2% (during 75%~99%)
±3% (during 50%~74%)
SpO
2
alarm:
Lower limit: 90%
D. Pulse Rate Parameter Specifications
Measuring range: 30bpm~240bpm
Accuracy: ±2bpm or ±2% (whichever is greater)
Pulse Rate alarm: Upper limit: 120bpm
Lower limit: 50bpm
E. Blood Perfusion Display
Range: 0~20%
F. Resistance to interference of surrounding light
The difference between the value measured in the condition of indoor natural light and that
of darkroom is less than ±1%.
G. Resistance to interference of man-made light
Values of SpO
2
and Pulse Rate can be accurately measured by pulse oxygen simulator.
H. Dimensions
66 mm (L) × 36 mm (W) × 33 mm (H)
Net Weight: 60g (including batteries)
I. Classification
The type of protection against electric shock: Internally powered equipment.
The degree of protection against electric shock: Type BF applied part.
The degree of protection against harmful ingress of liquids: Ordinary equipment without
protection against ingress of water.
Electro-Magnetic Compatibility: Group I, Class B
4. Accessories
A. A hanging cord
B. A pouch
C. A User’s Manual
ENGLISH
Summary of Contents for OXY-5
Page 55: ...55 2 1 1 1 pulsoximetro pulsoximetro...
Page 60: ...60 5 5 1 5 2 SpO2...
Page 61: ...61 5 3 1 2 6 A B C D 2 E F SpO2 HR SpO2 SpO2 BPM I 6...
Page 62: ...62 G PI H SpO2 7 C pulsoximetro D 20 C 55o C 10 100...
Page 63: ...63 SpO2 SpO2 1 2 SpO2 1 2 1 2 3 1 8 2 1 2 1 2 1 2 3 1 2 8...
Page 64: ...64 GIMA 12 GIMA GIMA software GIMA SHENZENCREATIVE P R C...
Page 65: ...65 ARABIC 2 1 1 1...
Page 66: ...66 ARABIC 1 2 MRI CT 1 3...
Page 67: ...67 ARABIC 60 2 HbO2 Hb O2 SpO2 SpO2 SpO2 SpO2 SpO2 2 1 SpO2...
Page 68: ...68 ARABIC 8 2 2 SpO2 2 3 10 40 20 55 35 80 10 100 70kPa 106kPa 50kPa 106kPa...
Page 71: ...71 ARABIC 0 20 Lower limit 50bpm 1 SpO2 66 36 33 60 BF I B 4...
Page 72: ...72 ARABIC 5 5 1 5 2 AAA ThedisplaySpO2 Thebar graphofpulsestrength Pulserate Power...
Page 73: ...73 ARABIC 5 3 1 2 6 2 SpO2 HR SpO2 SpO2 BPM...
Page 74: ...74 ARABIC 6 PI SpO2 7 20 55 10 100...
Page 75: ...75 ARABIC 8 1 2 1 2 SpO2 SpO2 1 2 1 2 SpO2 1 2 3 1 2 3 1 2 1 8 2...
Page 76: ...76 ARABIC 12 GIMA GIMA GIMA...