Specifications
Feature Specification
Meter Model Number
OGM-211
Measurement Range
0.6 - 33.3 mmol/L (10 – 600 mg/dL)
Result Calibration
Plasma-equivalent, calibrated by using YSI (Model
2300 STAT PLUS) Glucose Analyzer reference
instrument, which is traceable to NIST reference
standard.
Sample
Capillary, Venous, Arterial, Neonatal blood samples
Minimum Sample Size
0.6
μ
L
Test Time
5 seconds
On/Off Source
Two (2) CR 2032 3.0 V coin cell batteries
Battery Life
3,000 tests for glucose testing (not considering data
transfer)
Glucose Units of Measure
The meter is preset to either millimoles per liter
(mmol/L) or milligrams per deciliter (mg/dL)
depending on the standard of your country.
Memory
Up to 1000 records with time and date
Automatic shutoff
2 minutes after last action
Meter Size
90 mm × 60 mm × 16 mm
Display Size
43 mm × 40 mm
Weight
71.6 g (with batteries installed)
Operation Environmental
Conditions
Temperature: 5 - 45 ºC (41 – 113 ºF);
Relative Humidity: 10 - 90
%
(non-condensing);
Altitude:
≤
3048m
Meter storage and transportation
conditions:
Temperature:-20-50°C (-4 – 122 ºF);
Relative humidity: 10%-93% (non-condensing);
Air pressure: 500 hPa ~1060 hPa
Hematocrit Range
10 - 70
%
Data Port
9600 baud, 8 data bits, 1 stop bit, no parity