14
Press the
▼
or
▲
button to switch between turning the meal marker “On” and
“Off”. Press the
OK
button to confirm your selection.
7.
Audio Feature:
The meter comes with the meter audio feature enabled. The
meter will give one short beep when it is turned on, after sample detection and
when the result is ready. The meter will sound three short beeps to sound a
warning when an error has occurred. Please check the error number on the
display to confirm what kind of error has occurred.
Press the
▼
or
▲
button to switch between turning the meter beep “On” and “Off”.
Press the
OK
button to confirm your selection.
8.
Ketone Indicator:
The meter comes with the Ketone indicator feature
disabled. Press the
▼
or
▲
button to switch between turning the Ketone
indicator “On” and “Off”. Press the
OK
button to confirm your selection. When
the Ketone indicator is enabled, if the test result is higher than 16.7 mmol/L
(300 mg/dL), the symbol of "Ketone?" will appear on the display.
9.
Hyper Indicator:
The meter comes with the Hyper indicator feature disabled.
Press the
▼
or
▲
button to switch between turning the Hyper indicator “On”
and “Off”. Press the
OK
button to confirm your selection. When the Hyper
indicator is “Off”, pressing the
OK
button will go to the next Hypo indicator set
up. When the Hyper indicator is “On”, pressing the
OK
button will go to the
Hyper indicator level set up. At the Hyper level set up, press the
▼
or
▲
button to adjust the Hyper level then press the
OK
button to go to the Hypo
indicator set up.
OR