background image

b.  Alternatively, press “MEM” button in Clock Mode to display the stored 

results. The current memory bank will blink and the amount of results in this 

bank will be displayed. Press “START” button to change over to other bank. 

Confirm your selection by pressing “MEM” button. The current bank can 

also be confirmed automatically after 5 seconds with no operation.  

 

See picture 8-1.

c.  After selecting the memory bank, the LCD will display the average values in 

this bank. See picture 8-2&8-3&8-4. If no result stored, LCD will show zeros. 

See picture 8-5. Press “MEM” button to switch to next scene.

d.  When the average is displayed, press the “MEM” button, the most recent 

result will be displayed. See picture 8-6. Followed by, the blood pressure 

and pulse rate will be shown separately. Irregular heartbeat symbol (if any) 

will blink. Press “MEM” button again to review the next result. In this way, 

repeatedly pressing the “MEM” button displays the respective results  

measured previously.

e.  When displaying the stored results, the monitor will turn off automatically 

after 1 minute of no operation. You can also press the button “START”   

to turn off the monitor manually.

9.  Deleting measurements from the memory

When any result is displaying, keeping on pressing button “MEM” for three 

seconds, all results will be deleted. Press the button “MEM” or “START”,    

the monitor will turn off.

10. Assessing high blood pressure for adults

The following guidelines for assessing high blood pressure (without regard to 

age or gender) have been established by the World Health Organization (WHO).  

Please note that other factors (e.g. diabetes, obesity, smoking, etc.) need to be 

taken into consideration. Consult with your physician for accurate assessment, 

and never change your treatment by yourself.

Classification of blood pressure for adults

 

Note

: It is not intended to provide a basis of any type of rush toward  

emergency conditions/diagnosis based on the color scheme and that the color 

scheme is meant only to discriminate between the different levels of blood 

pressure. 

MAINTENANCE

1. 

Do not drop this monitor or subject it to strong impact.

2. 

Avoid high temperature and solarization. Do not immerse the monitor  

in water as this will result in damage to the monitor.

3.  If this monitor is stored near freezing, allow it to acclimate to room  

temperature before use.

4. 

Do not attempt to disassemble this monitor.

5.  If you do not use the monitor for a long time, please remove the  

batteries. 

6.  It is recommended the performance should be checked every 2 years   

or after repair. Please contact the service center.

7.  Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after 

moistened with water, diluted disinfectant alcohol, or diluted detergent.

8.  No component can be maintained by user in the monitor. The circuit    

diagrams, component part lists, descriptions, calibration instructions,   

or other information which will assist the user’s appropriately qualified  

 

technical personnel to repair those parts of equipment which are 

 

designated repairably can be supplied.

9.  The monitor can maintain the safety and performance characteristics 

 

for a minimum of 10,000 measurements or three years, and the cuff integrity 

is maintained after 1,000 open–close cycles of the closure.

11. Technical alarm description

The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if 

the determined blood pressure (systolic or diastolic) is outside the rated range 

specified in part SPECIFICACIONS. In this case, you should consult a physician 

or check if your operation violated the instructions.

The technical alarm condition (outside the rated range) is preset in the factory 

and cannot be adjusted or inactivated. This alarm condition is assigned as low 

priority according to IEC 60601-1-8.

The technical alarm is non-latching and need no reset. The signal displayed on 

LCD will disappear automatically after about 8 seconds.

13. Troubleshooting (2)

ELECTROMAGNETIC COMPATIBILITY INFORMATION

Table 1

For all ME EQUIPMENT and ME SYSTEMS

Symbol for “COMPILES WITH MDD93/42/EEC 

REQUIREMENTS”

Symbol for “DATE OF MANUFACTURE”

Symbol for “EUROPEAN REPRESENTATION”

Symbol for “SERIAL NUMBER”

Symbol for “KEEP DRY”

Symbol for “KEEP AWAY FROM SUNLIGHT”

Symbol for “PRODUCT CODE”

Symbol for “LOT NUMBER”

10. It is recommended the cuff should be disinfected 2 times every week    

if needed (For example, in hospital or in clinique). Wipe the inner side    

(the side contacts skin) of the cuff by a soft cloth squeezed after moistened 

with Ethyl alcohol (75-90%), then dry the cuff by airing.

EXPLANATION OF SYMBOLS ON UNIT

Symbol for “THE OPERATION GUIDE MUST BE READ” 

(The sign background colour: blue. The sign graphical symbol: white) 

Symbol for “WARNING”

Symbol for “TYPE BF APPLIED PARTS” 

(The cuff is type BF applied part)

Symbol for “ENVIRONMENT PROTECTION – Waste electrical 

products should not be disposed of with household waste. 

Please recycle where facilities exist. Check with your local Authority 

or retailer for recycling advice”.

Symbol for “MANUFACTURER”

Note

: Please consult a health care professional for interpretation of pressure 

measurements.

8.  Displaying stored results

a.  After the measurement, you can review the measurements in the current 

memory bank by pressing button “MEM”. Now the LCD displays the 

amount of the results in the current bank. 

 

Picture 8-1 

Picture 8-2 

Picture 8-3

 

 

Picture 8-4 

Picture 8-5 

Picture 8-6

M-32777-GB-Rev

.0.11.17

21

22

23

24

20

19

27

28

29

30

26

25

33

34

35

36

32

31

32777 / KD-5920

0197

ANDON HEALTH CO., LTD.

No. 3 JinPing Street, Ya An Road, Nankai District, 

Tianjin 300190, China

Made in China
Lotus Global Co., Ltd.

1 Four Seasons Terrace, West Drayton,

Middlesex, London, UB7 9GG, United Kingdom

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD Display shows 
battery symbol  

Low Battery

Change the batteries

LCD shows “Er 0”

Pressure system is unstable before 
measurement

Don’t move and try again

LCD shows “Er 1”

Fail to detect systolic pressure

LCD shows “Er 2”

Fail to detect diastolic pressure

LCD shows “Er 3”

Pneumatic system blocked or cuff 
is too tight during inflation

Apply the cuff correctly 
and try again

LCD shows “Er 4”

Pneumatic system leakage or cuff 
is too loose during inflation

LCD shows “Er 5”

Cuff pressure above 300mmHg

Measure again after five 
minutes. If the monitor is still 
abnormal, please contact the 
local distributor or the factory

LCD shows “Er 6”

More than 160 seconds with cuff 
pressure above 15 mmHg

LCD shows “Er 7”

Inner memory error

LCD shows “Er 8”

Device parameter checking error

LCD shows “Er A”

Pressure sensor parameter error

No response when you 
press button or load 
battery

Incorrect operation or strong 
electromagnetic interference

Take out batteries for five 
minutes, and then reinstall 
all batteries

Guidance and manufacture’s declaration - electromagnetic emissions

The 32777 is intended for use in the electromagnetic environment specified below.
The customer or the user of the 32777 should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions
CISPR 11

Group 1

The 32777 uses RF energy only for its internal 
function. Therefore, its RF emissions are very 
low and are not likely to cause any interference 
in nearby electronic equipment.

RF emissions
CISPR 11

Class B

The 32777 is suitable for use in all establishments 
other than domestic and those directly connected 
to the public low-voltage power supply network 
that supplies buildings used for domestic purposes.

Harmonic emissions
IEC 61000-3-2

Not applicable

Voltage fluctuations/
flicker emissions
IEC 61000-3-3

Not applicable

Table 2

For all ME EQUIPMENT and ME SYSTEMS

Table 3

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Table 4

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration - electromagnetic immunity

The 32777 is intended for use in the electromagnetic environment specified below. 
The customer or the user of the 32777 should assure that it is used in such an environment.

IMMUNITY TEST

IEC 60601test 
level

Compliance 
level

Electromagnetic environment 
- guidance

Electrostatic 
discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, 
concrete or ceramic tile. If floors 
are covered with synthetic 
material, the relative humidity 
should be at least 30%.

Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical commercial 
or hospital environment.

12. Troubleshooting (1)

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD Display shows 
abnormal result

The cuff position was not 
correct or it was not properly 
tightened

Apply the cuff correctly 
and try again

Body posture was not correct 
during testing

Review the “BODY POSTURE 
DURING MEASUREMENT” 
sections of the instructions 
and re-test

Speaking, arm or body 
movement, angry, excited or 
nervous during testing

Re-test when calm and 
without speaking or moving 
during the test

Irregular heartbeat 
(arrhythmia)

It is inappropriate for people 
with serious arrhythmia 
to use this Electronic 
Sphygmomanometer

Guidance and manufacturer’s declaration - electromagnetic immunity

The 32777 is intended for use in the electromagnetic environment specified below. 
The customer or the user of the 32777 should assure that it is used in such an environment.

IMMUNITY TEST

IEC 60601
test level

Compliance 
level

Electromagnetic environment - guidance

Radiated RF
IEC 61000-4-3

3 V/m 
80 MHz 
to 2.5 GHz

3 V/m

Portable and mobile RF communications 
equipment should be used no closer to any 
part of the 32777, including cables, than 
the recommended separation distance 
calculated from the equation applicable to 
the frequency of the transmitter.

Recommended separation distance

Where 

P

 is the maximum output power 

rating of the transmitter in watts (W) 
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (m).
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site 
survey,

a

 should be less than the compliance 

level in each frequency range.

b

Interference may occur in the 
vicinity of equipment marked 
with the following symbol:

NOTE 1 

At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 

cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due 

to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field 

strength in the location in which the 32777 is used exceeds the applicable RF compliance level 

above, the 32777 should be observed to verify normal operation. If abnormal performance 

is observed, additional measures may be necessary, such as re-orienting or relocating the 32777.

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

= 2.3 √ 

P

800 MHz to 2,5 GHz

Disposal

: The product must not be disposed of along with other domestic 

waste. The users must dispose of this equipment by bringing it to a 

specific recycling point for electric and electronic equipment. For further 

information on recycling points contact the local authorities, the local 

recycling center or the shop where the product was purchased. If the 

equipment is not disposed of correctly, fines or penalties may be applied

in accordance with the national legislation and regulations.

GIMA WARRANTY CONDITIONS

Congratulations for purchasing a GIMA product. This product meets high qualitative 

standards both as regards the material and the production. 

The warranty is valid for 12 months from the date of supply of GIMA.

During the period of validity of the warranty, GIMA will repair and/or replace 

free of charge all the defected parts due to production reasons. 

Labor costs and personnel traveling expenses and packaging not included.

All components subject to wear are not included in the warranty.

The repair or replacement performed during the warranty period shall not extend 

the warranty. The warranty is void in the following cases: repairs performed by 

unauthorized personnel or with non-original spare parts, defects caused by negligence 

or incorrect use. GIMA cannot be held responsible for malfunctioning on electronic 

devices or software due to outside agents such as: voltage changes, electro-magnetic 

fields, radio interferences, etc. 

The warranty is void if the above regulations are not observed and if the serial code 

(if available) has been removed, cancelled or changed.

The defected products must be returned only to the dealer the product was purchased 

from. Products sent to GIMA will be rejected.

Recommended separation distances between portable and mobile RF communications 
equipment and the 32777

The 32777 is intended for use in an electromagnetic environment in which radiated RF disturbances 
are controlled. The customer or the user of the 32777 can help prevent electromagnetic interference 
by maintaining a minimum distance between portable and mobile RF communications equipment 
(transmitters) and the 32777 as recommended below, according to the maximum output power 
of the communications equipment.

Rated maximum 
output power 
of transmitter W

Separation distance according to frequency of transmitter m

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2,5 GHz

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation 

distance 

d

 in meters (m) can be determined using the equation applicable to the frequency of the 

transmitter, where 

P

 is the maximum output power rating of the transmitter in watts (W) according 

to the transmitter manufacturer.

NOTE 1

 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

= 1.2 √ 

P

= 1.2 √ 

P

= 2.3 √ 

P

0197

SN

WHO/ISH Definitions and classification of blood pressure levels

80

85

90

100

110

120

130

140

150

160

Severe Hypertension

Moderate Hypertension

Mild Hypertension

High-normal BP

Normal BP

Diastolic 

(mmHg)

Sistolic 

(mmHg)

BLOOD PRESSURE 
CLASSIFICATION

SBP 

mmHg

DBP

 

mmHg

Color

Indicator

Optimal 

<120

<80

green

Normal

120-129

80-84

green

High-Normal

130-139

85-89

green

Grade 1 Hypertension

140-159

90-99

yellow

Grade 2 Hypertension

160-179

100-109

orange

Grade 3 Hypertension

≥180

≥110

red

Reviews: