3.4 Data Uploading
1. When the wireless transmission function is on, the monitor can communicate with a host
device such as a PC, smart phone or other wireless enabled devices for real-time data tran-
smission.
2. When connected to a USB cable, the history data can be uploaded to a PC for viewing and
management.
Chapter 4
TECHNICAL SPECIFICATIONS
4.1 Blood Pressure Measurement
1. Technique: Oscillometric
2. Pressure measuring range: 0mmHg~300mmHg
3. Cuff inflation time:<20 seconds (typical adult cuff)
4. Overpressure protection limit
Adult: 300mmHg (39.9kPa);
Child: 240mmHg;
Infant: 150mmHg
5. Blood pressure measurement range:
SYST: 30mmHg~270mmHg
DIAS: 20mmHg~235mmHg
MAP: 10mmHg~220mmHg
6. Blood pressure measuring accuracy:
Maximal mean difference: ≤ ±5mmHg (0.67kPa)
Maximal standard deviation: ≤8mmHg (1.067kPa)