same place for over two hours, change the measurement site periodically and when neces-
sary.
Do NOT place the SpO
2
probe on a finger with edema or fragile tissue.
When the ambient temperature is over 35°C, please change the measuring site every two
hours; when the ambient temperature is over 37°C, please do NOT use the SpO
2
sensor, as
using in high temperatures can cause burns.
Do NOT put the SpO
2
probe and pressure cuff on the same limb, otherwise the blood pres-
sure measurement may affect the SpO
2
measurement.
The device is calibrated to display functional oxygen saturation.
Do NOT allow the sensor cable to twist or bend.
Check the SpO
2
sensor and cable before use. Do NOT use a damaged SpO
2
sensor.
When the temperature of the SpO
2
sensor is abnormal, do not use it further.
Remove nail polisher or other cosmetic products from the fingernail.
The fingernail should be of normal length.
The SpO
2
sensor cannot be immersed into water, liquid or cleanser.
The SpO
2
sensor can be repeatedly used. Please clean and disinfect before reuse.
The SpO
2
sensor can be repeatedly used. Please clean and disinfect before reuse.
Anemia or low hemoglobin concentrations, intravascular dyes, carboxyhemoglobin, methe-
moglobin, and dysfunctional hemoglobin may effect the SpO
2
accuracy. If the patient has
such situation, do not rely on the measured result for diagnostic decision, and it's recommen-
ded for the patient to consult with the doctor.
☞
PORT1 or PORT2
can be connected with either temperature probe, smart SpO
2
probe, or
Easy ECG Monitor, but not any other devices or probes. Do NOT connect two probes or
devices with the same type (e.g. two temperature probes, or two smart SpO
2
probes, or two
Easy ECG Monitors) to both PORT1 and PORT2.